Comparison between the Datex-Ohmeda M-NMT module and a force-displacement transducer for monitoring neuromuscular blockade.

C Motamed1, K Kirov, X Combes, P Duvaldestin.   

Abstract

BACKGROUND AND
OBJECTIVE: The Datex-Ohmeda neuromuscular transmission module (M-NMT) is a new monitor that is part of the AS/3 anaesthesia monitor. It incorporates a mechanosensor, which is a piezoelectric polymer attached to the hand. The module was compared with a force transducer in 30 patients requiring neuromuscular blockade.
METHODS: Anaesthesia was induced with fentanyl and propofol, and tracheal intubation was performed without muscle relaxants. Neuromuscular blockade was assessed by the test module on one arm and the force transducer on the other arm. When the response to train-of-four stimulation had been stable in both arms for 3 min, rocuronium 0.2 mg kg(-1) was injected intravenously. During recovery from blockade, the train-of-four ratio was measured in 15 patients, and the ratio of response to double-burst stimulation in the other 15 patients. Data analysis was by difference plots.
RESULTS: Both devices had acceptable coefficients of repeatability. The M-NMT module was biased by + 1.3% (upper limit of agreement 14.2%, lower limit -12.9%) for the recovery of the train-of-four ratio, and by + 1.09% (17%, -16%) for the recovery of double-burst stimulation ratio.
CONCLUSIONS: The Datex-Ohmeda M-NMT gives measurements that are repeatable and gives good enough correspondence with a force transducer that it can be used clinically to assess recovery of neuromuscular blockade, but the limits of agreement rule out research applications.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12803264     DOI: 10.1017/s0265021503000735

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  6 in total

1.  Computerized recording of neuromuscular monitoring and the risk of residual paralysis at the time of extubation.

Authors:  Cyrus Motamed; Jean-Louis Bourgain
Journal:  J Clin Monit Comput       Date:  2008-08-07       Impact factor: 2.502

2.  Survey of muscle relaxant effects management with a kinemyographic-based data archiving system: a retrospective quantitative and contextual quality control approach.

Authors:  Cyrus Motamed; Jean Louis Bourgain; Alain D'Hollander
Journal:  J Clin Monit Comput       Date:  2013-07-10       Impact factor: 2.502

3.  Comparison of train-of-four ratios measured with Datex-Ohmeda's M-NMT MechanoSensor™ and M-NMT ElectroSensor™.

Authors:  Jarno Salminen; Mark van Gils; Markku Paloheimo; Arvi Yli-Hankala
Journal:  J Clin Monit Comput       Date:  2015-07-08       Impact factor: 2.502

4.  Lateral spread response monitoring during microvascular decompression for hemifacial spasm. Comparison of two targets of partial neuromuscular blockade.

Authors:  Y H Chung; W H Kim; J J Lee; S-I Yang; S H Lim; D W Seo; K Park; I S Chung
Journal:  Anaesthesist       Date:  2014-02-07       Impact factor: 1.041

Review 5.  [Algorithm-based preventive strategies for avoidance of residual neuromuscular blocks].

Authors:  C Unterbuchner; K Ehehalt; B Graf
Journal:  Anaesthesist       Date:  2019-11       Impact factor: 1.041

6.  Comparison of Train of Four Measurements with Kinemyography NMT DATEX and Accelerography TOFscan.

Authors:  Cyrus Motamed; Migena Demiri; Nora Colergrave
Journal:  Med Sci (Basel)       Date:  2021-03-29
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.