Commissioning and clinical implementation of a sliding gantry CT scanner installed in an existing treatment room and early clinical experience for precise tumor localization.

Chee-Wai Cheng1, James Wong, Lisa Grimm, Michael Chow, Minoru Uematsu, Albert Fung.   

Abstract

The primary objective of the present study is to demonstrate that a unique computed tomography (CT)-linear accelerator combination can be used to reduce uncertainties caused by organ motion and setup inaccuracy. The acceptance, commissioning, and clinical implementation of a sliding gantry CT scanner installed in an existing linear accelerator room are reported in this paper. A Siemens CT scanner was installed directly opposite to an existing accelerator. The scanner is movable on a pair of horizontal rails mounted parallel to the longitudinal axis of the treatment couch replaced with a carbon fiber tabletop. Acceptance and commissioning of the CT scanner were verified with phantom studies. For clinical implementation, quality assurance (QA) procedures have been instituted to ensure the integrity of the CT gantry axis alignment and the accuracy of its movement using a phantom designed in house. A clinical example employing the CT-Linac combination to correct the isocenter positioning caused by organ motion and setup inaccuracy was presented for a prostate irradiation. Dose calculations were performed to study the effects on tumor coverage without the adjustments of the isocenter. A summary of the isocenter adjustments for the first 30 patients is also presented. The geometric accuracy of the CT scanner is < or =1 mm. An isocenter deviation of > or =2 mm from the original plan can be detected. For the clinical example of a prostate patient, the average movement of the prostate gland was found to be approximately 3mm in the anterior-posterior (AP/PA) direction and 5 mm in the cephalic-caudal direction. Variations in the isocenter position may result in underdosage of the PTV if correction is not made for the change in the isocenter position. Our experience with the first 30 patients indicates that while the left-right adjustment of the isocenter is minimal, in the AP/PA direction, about 33% of treatments required an adjustment of 3-5 mm, and about 18% required a 5.1-mm to 10-mm adjustment. In the caudal-cephalic direction, about 26% required an adjustment of 3-5 mm, and 8% required a 5.1-mm to 10-mm adjustment. Retrofitting a CT scanner in an existing linear accelerator room requires careful planning and well-coordinated efforts from all personnel involved. Special QA procedures are needed to ensure the mechanical integrity and imaging accuracy of the CT scanner. A CT scan of the patient prior to irradiation provides valuable information on organ motion. Any deviations from treatment plan can be corrected before dose delivery. Significant deviation from the planning isocenter may occur due to daily variations in the rectal filling. The CT-Linac combination has significant implications for the treatment of prostate cancer.

Entities:  

Mesh:

Year:  2003        PMID: 12796613     DOI: 10.1097/01.COC.0000072509.66808.2C

Source DB:  PubMed          Journal:  Am J Clin Oncol        ISSN: 0277-3732            Impact factor:   2.339


  5 in total

Review 1.  Radiation oncology in the era of precision medicine.

Authors:  Michael Baumann; Mechthild Krause; Jens Overgaard; Jürgen Debus; Søren M Bentzen; Juliane Daartz; Christian Richter; Daniel Zips; Thomas Bortfeld
Journal:  Nat Rev Cancer       Date:  2016-03-18       Impact factor: 60.716

2.  Current concepts on imaging in radiotherapy.

Authors:  Michela Lecchi; Piero Fossati; Federica Elisei; Roberto Orecchia; Giovanni Lucignani
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-31       Impact factor: 9.236

Review 3.  Imaging in radiation oncology: a perspective.

Authors:  Laura A Dawson; Cynthia Ménard
Journal:  Oncologist       Date:  2010

4.  Significance of image guidance to clinical outcomes for localized prostate cancer.

Authors:  Qiuzi Zhong; Hong Gao; Gaofeng Li; Xia Xiu; Qinhong Wu; Ming Li; Yonggang Xu
Journal:  Biomed Res Int       Date:  2014-07-13       Impact factor: 3.411

5.  Influence of acquisition parameters on MV-CBCT image quality.

Authors:  Olivier Gayou
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.