Apparent P wave undersensing in a DDD pacemaker post exercise.

B M van Gelder1, R van Mechelen, K den Dulk, M I el Gamal.   

Abstract

Wenckebach behavior of DDD pacemakers occurring when the P-P interval varies between the programmed upper rate interval and the total atrial refractory period is symmetrical in a sense that the pacemaker response during atrial rate acceleration is similar to the pacemaker response during atrial rate deceleration. This phenomenon can be observed in all patients with persistent AV block in whom a DDD pacemaker is implanted, during exercise testing when the spontaneous atrial rate exceeds the selected upper rate, i.e., the programmed upper rate interval. However, this phenomenon will not be observed in all patients with intermittent intact AV conduction during exercise. In this case report we describe a patient who showed an asymmetrical response during a bicycle exercise test. There was 1:1 atrial sensing ventricular pacing until the atrial rate exceeded the upper rate of 140 ppm, while atrial sensing was restored during recovery when the conducted sinus rhythm had decreased to 105 beats/min.

Entities:  

Mesh:

Year:  1992        PMID: 1279530     DOI: 10.1111/j.1540-8159.1992.tb02950.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Disappearance of orthostatic hypotension after pacemaker implantation in a patient with a long PR interval.

Authors:  Elia De Maria; Pier Luigi Fontana; Stefano Cappelli
Journal:  Indian Heart J       Date:  2012-12-26

2.  First-degree atrioventricular block. Clinical manifestations, indications for pacing, pacemaker management & consequences during cardiac resynchronization.

Authors:  S Serge Barold; Arzu Ilercil; Fabio Leonelli; Bengt Herweg
Journal:  J Interv Card Electrophysiol       Date:  2007-03-02       Impact factor: 1.900

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.