Literature DB >> 12794400

A discard volume of twice the deadspace ensures clinically accurate arterial blood gases and electrolytes and prevents unnecessary blood loss.

Claire M Rickard1, Bronwyn A Couchman, Sharon J Schmidt, Alexandra Dank, David M Purdie.   

Abstract

OBJECTIVE: To determine the blood discard volume, as a multiple of deadspace, that is required for accurate arterial blood gas and electrolyte testing from arterial catheters.
DESIGN: Prospective, controlled, crossover trial.
SETTING: Eighteen-bed intensive care unit of a metropolitan teaching hospital. PATIENTS: A total of 84 critically ill patients with a 20-gauge, radial arterial cannulae, pressure monitoring transducer set, and stable oxygenation.
INTERVENTIONS: System deadspace (priming volume from sampling port to catheter tip) was established. Patients had six 0.5-mL arterial blood samples taken sequentially in random order using discard volumes of 1, 1.5, 2, 2.3, and 3.6 times the deadspace (experimental values) and 5.5 times the deadspace (control).
MEASUREMENTS AND MAIN RESULTS: Samples were analyzed for Pao(2), Sao(2), pH, Paco(2), HCO(3)-, Na+, and K+. We performed repeated-measures analysis of variance with post hoc linear contrasts and compared mean experimental and control values. The smallest discard volumes that provided measurements that were statistically equal to control were twice the deadspace (Pao(2), p =.563; Sao(2), p =.371) and 3.6 times the deadspace (pH, p =.107; Paco(2), p=.519; HCO(3)-, p =.10). All discard volumes tested provided results that were statistically different from control for Na+ (p <.003) and K+ (p <.001).
CONCLUSIONS: Many results were statistically different from control, although the actual discrepancies were very small. At clinically relevant levels of measurement, there was minimal variation between values obtained after a discard volume of twice the deadspace and control values. The level of error was clinically acceptable and within or close to the precision limits of the blood gas analyzer. Slight fluctuation in patient variables during sampling could also have contributed to the error. A blood discard volume of twice the deadspace is recommended for all variables. This will provide clinically accurate results and avoid the deleterious effects of unnecessary blood loss.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12794400     DOI: 10.1097/01.CCM.0000063448.98777.EF

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  3 in total

1.  A Standard Push-Pull Protocol for Waste-Free Sampling in the Pediatric Intensive Care Unit.

Authors:  Clare McBride; Suzan Miller-Hoover; James A Proudfoot
Journal:  J Infus Nurs       Date:  2018 May/Jun

2.  Inaccuracy of Venous Point-of-Care Glucose Measurements in Critically Ill Patients: A Cross-Sectional Study.

Authors:  Adriano José Pereira; Thiago Domingos Corrêa; Francisca Pereira de Almeida; Rodrigo Octávio Deliberato; Michelle dos Santos Lobato; Nelson Akamine; Eliézer Silva; Alexandre Biasi Cavalcanti
Journal:  PLoS One       Date:  2015-06-12       Impact factor: 3.240

3.  Repeatability of blood gas parameters, PCO2 gap, and PCO2 gap to arterial-to-venous oxygen content difference in critically ill adult patients.

Authors:  Jihad Mallat; Ali Lazkani; Malcolm Lemyze; Florent Pepy; Mehdi Meddour; Gaëlle Gasan; Johanna Temime; Nicolas Vangrunderbeeck; Laurent Tronchon; Didier Thevenin
Journal:  Medicine (Baltimore)       Date:  2015-01       Impact factor: 1.889

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.