Literature DB >> 12791110

Measurement of cerebral oxygenation state in anaesthetized children using the INVOS 5100 cerebral oximeter.

Alexander Dullenkopf1, Bernhard Frey, Oskar Baenziger, Andreas Gerber, Markus Weiss.   

Abstract

BACKGROUND: Near-infrared spectroscopy is a developing technique for monitoring cerebral oxygenation during anaesthesia. The aim of this study was to evaluate absolute values of cerebral oxygenation during stable anaesthesia conditions in otherwise healthy children using the recently introduced INVOS 5100 cerebral oximeter with a paediatric and adult sensor and to compare them with values obtained from the NIRO 300 oximeter.
METHODS: Thirty paediatric surgical patients (aged 0.23-15.97 years) were studied during general anaesthesia with tracheal intubation and controlled ventilation. Comparative measurements of cerebral oxygenation were performed on the forehead with two probes within 10 min under stable cardiorespiratory and anaesthesia conditions. Cerebral oxygenation values (rSO2) obtained from the paediatric and adult INVOS 5100 sensors were compared with the tissue oxygenation index (TOI) obtained from the NIRO 300 cerebral oximeter using 4- and 5-cm emitter-detector separation.
RESULTS: Cerebral rSO2 values and the TOI values both showed a large range of cerebral oxygenation in the children studied (rSO2: 59-95%, TOI: 48-85%). Cerebral rSO2 values measured by the INVOS 5100, particularly with the paediatric sensor, were significantly higher than the TOI values obtained from the NIRO 300 (P < 0.0001). Agreement between the INVOS and NIRO oximeter was poor.
CONCLUSION: The large range and the poor agreement of cerebral oxygenation values between the two oximeters makes it difficult to define a normal value. Cerebral oxygenation readings by these monitors, based on one single point measurement during anaesthesia, should be viewed with caution. Actually, there may be little indication for routine use of such monitoring during general anaesthesia.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12791110     DOI: 10.1046/j.1460-9592.2003.01111.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  25 in total

1.  A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section.

Authors:  Simon Hyttel-Sorensen; Trine Witzner Hessel; Amalia la Cour; Gorm Greisen
Journal:  Biomed Opt Express       Date:  2014-09-18       Impact factor: 3.732

2.  Effects of maternal supplementary oxygen on the newborn for elective cesarean deliveries under spinal anesthesia.

Authors:  Berrin Gunaydin; Tuncay Nas; Aydan Biri; Esin Koc; Ayfer Koc; Kevin McCusker
Journal:  J Anesth       Date:  2011-03-20       Impact factor: 2.078

3.  Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom.

Authors:  Simon Hyttel-Sorensen; Stefan Kleiser; Martin Wolf; Gorm Greisen
Journal:  Biomed Opt Express       Date:  2013-08-14       Impact factor: 3.732

Review 4.  In vivo validation of cerebral near-infrared spectroscopy: a review.

Authors:  Amalie la Cour; Gorm Greisen; Simon Hyttel-Sorensen
Journal:  Neurophotonics       Date:  2018-11-27       Impact factor: 3.593

5.  Quantifying the effect of adipose tissue in muscle oximetry by near infrared spectroscopy.

Authors:  Nassim Nasseri; Stefan Kleiser; Daniel Ostojic; Tanja Karen; Martin Wolf
Journal:  Biomed Opt Express       Date:  2016-10-17       Impact factor: 3.732

6.  Comparison of tissue oximeters on a liquid phantom with adjustable optical properties.

Authors:  S Kleiser; N Nasseri; B Andresen; G Greisen; M Wolf
Journal:  Biomed Opt Express       Date:  2016-07-11       Impact factor: 3.732

7.  Comparison of tissue oximeters on a liquid phantom with adjustable optical properties: an extension.

Authors:  S Kleiser; D Ostojic; B Andresen; N Nasseri; H Isler; F Scholkmann; T Karen; G Greisen; M Wolf
Journal:  Biomed Opt Express       Date:  2017-12-05       Impact factor: 3.732

8.  Cerebral oximetry in preterm infants: an agenda for research with a clear clinical goal.

Authors:  Gorm Greisen; Bjørn Andresen; Anne Mette Plomgaard; Simon Hyttel-Sørensen
Journal:  Neurophotonics       Date:  2016-04-25       Impact factor: 3.593

9.  Correlation between Regional Cerebral Saturation and Invasive Cardiac Index Monitoring after Heart Transplantation Surgery.

Authors:  Awni M Al-Subu; Christoph P Hornik; Ira M Cheifetz; Andrew J Lodge; George Ofori-Amanfo
Journal:  J Pediatr Intensive Care       Date:  2018-06-11

10.  Cerebral oximetry during infant cardiac surgery: evaluation and relationship to early postoperative outcome.

Authors:  Barry D Kussman; David Wypij; James A DiNardo; Jane W Newburger; John E Mayer; Pedro J del Nido; Emile A Bacha; Frank Pigula; Ellen McGrath; Peter C Laussen
Journal:  Anesth Analg       Date:  2009-04       Impact factor: 5.108

View more