Literature DB >> 12788196

(Non)-migration of radiopaque markers used for on-line localization of the prostate with an electronic portal imaging device.

Jean Pouliot1, Michele Aubin, Katja M Langen, Yu-Ming Liu, Barby Pickett, Katsuto Shinohara, Mack Roach.   

Abstract

PURPOSE: Radiopaque gold markers can be implanted in the prostate to visualize its position on portal images during radiation therapy. This procedure assumes that the markers do not move within the prostate. In this work we test this assumptiom. METHODS AND MATERIALS: Three markers were implanted transrectally in the prostate of patients undergoing external radiation therapy. An orthogonal pair of portal images was acquired periodically throughout the course of radiation therapy with an a-Si electronic portal imaging device (EPID). The marker coordinates were determined, and the distances between the implanted markers were recorded. The distance time trend is used to evaluate the magnitude of marker migration.
RESULTS: The average standard deviation (SD) of the distances between markers was 1.3 mm (range 0.44 to 3.04 mm). Three of the 11 patients show a SD larger than 2 mm. For these patients, all three distances show a simultaneous reduction with time, compatible with a shrinking of the prostate. All had been treated with neoadjuvant hormone therapy. For 1 of the 3 patients, this reduction in volume was confirmed with a repeat computed tomographic scan.
CONCLUSION: None of the 33 markers studied migrated significantly. The implantation of three radiopaque gold markers enables accurate and precise on-line verification of the prostate position during external beam radiation therapy. The use of three markers provides a tool to monitor prostate position and volume changes that can occur over time due to hormone or radiation therapy.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12788196     DOI: 10.1016/s0360-3016(03)00267-0

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  18 in total

1.  Intrafractional motion of the prostate during hypofractionated radiotherapy.

Authors:  Yaoqin Xie; David Djajaputra; Christopher R King; Sabbir Hossain; Lijun Ma; Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-09-01       Impact factor: 7.038

2.  Fast internal marker tracking algorithm for onboard MV and kV imaging systems.

Authors:  W Mao; R D Wiersma; L Xing
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 3.  Fiducial marker guided prostate radiotherapy: a review.

Authors:  Angela G M O'Neill; Suneil Jain; Alan R Hounsell; Joe M O'Sullivan
Journal:  Br J Radiol       Date:  2016-09-23       Impact factor: 3.039

4.  Analysis of fiducial markers used for on-line verification in the external-beam radiotherapy of patients with cranial tumours.

Authors:  R Ma Cañón; I Azinovic; M Lobato; J Navarro; J Rebollo
Journal:  Clin Transl Oncol       Date:  2007-08       Impact factor: 3.405

5.  A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging.

Authors:  Weihua Mao; Nadeem Riaz; Louis Lee; Rodney Wiersma; Lei Xing
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

6.  Comparison of daily couch shifts using MVCT (TomoTherapy) and B-mode ultrasound (BAT System) during prostate radiotherapy.

Authors:  Steven H Lin; Elizabeth Sugar; Terrance Teslow; Todd McNutt; Habeeb Saleh; Danny Y Song
Journal:  Technol Cancer Res Treat       Date:  2008-08

7.  Six-Dimensional Correction of Intra-Fractional Prostate Motion with CyberKnife Stereotactic Body Radiation Therapy.

Authors:  Siyuan Lei; Nathaniel Piel; Eric K Oermann; Viola Chen; Andrew W Ju; Kedar N Dahal; Heather N Hanscom; Joy S Kim; Xia Yu; Guowei Zhang; Brian T Collins; Reena Jha; Anatoly Dritschilo; Simeng Suy; Sean P Collins
Journal:  Front Oncol       Date:  2011-12-08       Impact factor: 6.244

Review 8.  Is there an optimal management for localized prostate cancer?

Authors:  Jaspreet Singh; Edouard J Trabulsi; Leonard G Gomella
Journal:  Clin Interv Aging       Date:  2010-08-09       Impact factor: 4.458

9.  Third party EPID with IGRT capability retrofitted onto an existing medical linear accelerator.

Authors:  D O Odero; D S Shimm
Journal:  Biomed Imaging Interv J       Date:  2009-07-01

10.  Computed tomography based evaluation of prostatic fiducial marker migration between the periods of insertion and simulation.

Authors:  Taner Arpacı; Gamze Uğurluer; Emine Burçin İspir; Alper Eken; Tuğana Akbaş; Meltem Serin
Journal:  Turk J Urol       Date:  2017-12-01