Intraoperative high-dose-rate 192Ir radical implant in early breast cancer: a quality assurance and dosimetry study.

S Sowmya Narayanan1, Vineeta Goel, Rajiv Sarin, Rakesh Jalali, Shyam Kishore Shrivastava, Deepak Dattatray Deshpande, Ketayun Ardeshir Dinshaw.   

Abstract

PURPOSE: To evaluate the variability in catheter length, geometry, and dosimetric parameters of radical intraoperative high-dose-rate breasts implant during 7-11 days. METHODS AND MATERIALS: Simulator X-rays, CT scans, and dosimetric studies were repeated on alternate days in 14 consecutive patients treated with radical intraoperative two- or three-plane nylon catheter high-dose-rate implant (34 Gy in 10 fractions within 5 days).
RESULTS: Significant variation was found in catheter length, but no major change was noted in implant geometry, homogeneity, or inhomogeneity indexes. A variation in length of >5 mm in one or more catheters was seen in all patients and >10 mm in 11 patients at any time during the implant. Of the 171 catheters in 14 patients, 100 (58%) and 38 (22%) showed a variation of >5 mm or >10 mm, respectively. The variation of >10 mm was reduced from 32% of catheters in the first 5 patients to 17% in the subsequent 9 patients (p = 0.028). Rigid catheter fixation might reduce length variation but may cause skin necrosis if the expanding cavity indents the skin for a long period against the fixation device. A dose homogeneity index of 0.90 (range 0.85-0.92) and dose nonuniformity ratio of 0.20 (range 0.12-0.25) were satisfactory.
CONCLUSION: The catheter fixation and exit catheter length should be measured daily and if the implant is in situ for more than a few days, orthogonal X-rays and, if indicated, dosimetry should be repeated at least once.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12788174     DOI: 10.1016/s0360-3016(03)00015-4

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Quality assessment of interstitial implants in high- dose- rate brachytherapy after lumpectomy in patients of early stage breast cancer.

Authors:  Samrat Dutta; Suresh C Sharma; Rajinder Singh; Rakesh Kapoor; Arun S Oinam; Anoop Bhardwaj; Pradeep Goswami
Journal:  Indian J Surg Oncol       Date:  2011-03-24

Review 2.  Breast cancer: An overview of published Indian data.

Authors:  Bharath Rangarajan; Tanuja Shet; Tabassum Wadasadawala; Nita S Nair; R Madhu Sairam; Sachin S Hingmire; Jyoti Bajpai
Journal:  South Asian J Cancer       Date:  2016 Jul-Sep
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.