The purpose of this study was to evaluate differences in fit and fill between an anatomic femoral component and a straight tapered femoral component, both of which were designed for proximal fit and fill using the preoperative planning workstation of the ROBODOC system (ORTHODOC). Anatomic Hip (Zimmer) and VerSys Taper (Zimmer) femoral components were each virtually implanted into 50 femora (25 dysplastic femora and 25 anatomically normal femora) using the ORTHODOC workstation. The fit and fill of the femoral components were measured on cross-sectional images. The VerSys Taper femoral components showed significantly better fit and fill than the Anatomic Hip femoral components at the lower corner of the femoral neck cut and the middle of the femoral component in both the dysplastic femora and the anatomically normal femora. The Anatomic Hip femoral components showed significantly better fit and fill than the VerSys Taper femoral components 1 cm proximal from the femoral component tip in both dysplastic femora and anatomically normal femora. There was no significant difference in fit and fill between the two types of femoral component at the center of the lesser trochanter or 1 cm distal from the center of the lesser trochanter in either dysplastic femora or anatomically normal femora. Overall, VerSys Taper femoral components appear to provide better proximal fit and fill than Anatomic Hip femoral components in both dysplastic and anatomically normal femora.

Entities:  

Mesh:

Year:  2003        PMID: 12768478     DOI: 10.1007/s10776-002-0637-0

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  4 in total

1.  Usefullness of three-dimensional templating software to quantify the contact state between implant and femur in total hip arthroplasty.

Authors:  Daisuke Inoue; Tamon Kabata; Toru Maeda; Yoshitomo Kajino; Kenji Fujita; Kazuhiro Hasegawa; Takashi Yamamoto; Tomoharu Takagi; Takaaki Ohmori; Hiroyuki Tsuchiya
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-11-05

2.  A comparison between robotic-assisted and manual implantation of cementless total hip arthroplasty.

Authors:  Nobuo Nakamura; Nobuhiko Sugano; Takashi Nishii; Akihiro Kakimoto; Hidenobu Miki
Journal:  Clin Orthop Relat Res       Date:  2009-11-05       Impact factor: 4.176

3.  Does Robotic Milling For Stem Implantation in Cementless THA Result in Improved Outcomes Scores or Survivorship Compared with Hand Rasping? Results of a Randomized Trial at 10 Years.

Authors:  Nobuo Nakamura; Nobuhiko Sugano; Takashi Sakai; Ichiro Nakahara
Journal:  Clin Orthop Relat Res       Date:  2018-11       Impact factor: 4.176

4.  The correlation between clinical radiological outcome and contact state of implant and femur using three-dimensional templating software in cementless total hip arthroplasty.

Authors:  Daisuke Inoue; Tamon Kabata; Toru Maeda; Yoshitomo Kajino; Takashi Yamamoto; Tomoharu Takagi; Takaaki Ohmori; Hiroyuki Tsuchiya
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-06-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.