| Literature DB >> 12763530 |
Hohyun Kim1, Jeong-Yeon Yoo, Sang Beom Han, Hee Joo Lee, Kyung Ryul Lee.
Abstract
A sensitive and selective liquid chromatographic method coupled with tandem mass spectrometry (LC-MS/MS) was developed for the quantification of ambroxol in human plasma. Domperidone was used as internal standard, with plasma samples extracted using diethyl ether under basic condition. A centrifuged upper layer was then evaporated and reconstituted with 200 microl methanol. The reconstituted samples were injected into a C(18) XTerra MS column (2.1 x 30 mm) with 3.5 microm particle size. The analytical column lasted for at least 600 injections. The mobile phase was composed of 20 mM ammonium acetate in 90% acetonitrile (pH 8.8), with flow rate at 250 microl/min. The mass spectrometer was operated in positive ion mode using turbo electrospray ionization. Nitrogen was used as the nebulizer, curtain, collision, and auxiliary gases. Using MS/MS with multiple reaction monitoring (MRM) mode, ambroxol was detected without severe interferences from plasma matrix. Ambroxol produced a protonated precursor ion ([M+H](+)) at m/z 379 and a corresponding product ion at m/z 264. And internal standard (domperidone) produced a protonated precursor ion ([M+H](+)) at m/z 426 and a corresponding product ion at m/z 174. Detection of ambroxol in human plasma was accurate and precise, with quantification limit at 0.2 ng/ml. This method has been successfully applied to a study of ambroxol in human specimens.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12763530 DOI: 10.1016/s0731-7085(03)00059-1
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935