Development and testing of a pelvic goniometer designed to measure pelvic tilt and hip flexion.

Stephen Sprigle1, Nannette Flinn, Mary Wootten, Stephanie McCorry.   

Abstract

OBJECTIVE: To determine the reliability and validity of a pelvic goniometer designed to measure the pelvic tilt and hip flexion during seated posture.
BACKGROUND: Assessment of the seated posture requires measurement of the pelvis and hip. Determining accurate pelvic tilt and hip flexion angles during sitting is often difficult using standard techniques. A pelvic goniometer has been designed to measure pelvic tilt and hip flexion angle of persons in a seated posture.
METHODS: VALIDATION of the pelvic goniometer was done radiographically. Ten male volunteers sat in three postures--erect, forward or anterior tilt, and posterior tilt. Pelvic tilt and hip angle were recorded using radiographs and the pelvic goniometer. Reliability of pelvic and conventional goniometers was done using seated nondisabled subjects with physical therapists performing measurements. VALIDATION: the average differences and correlation between the pelvic goniometer and radiographic measures were as follows--pelvic tilt: -4.9 degrees, 0.93; hip angle 1.2 degrees, 0.81. Reliability: average range of hip angle across three measures was about 3 degrees for both goniometers.
CONCLUSIONS: The data indicate that the pelvic goniometer has utility in measuring pelvic tilt and hip angle, especially within the seated posture. Because it measures both pelvic tilt and hip angle, the pelvic goniometer has an advantage over conventional goniometers that only measure the latter. RELEVANCE: A valid and reliable tool that measures pelvic tilt and hip angle of persons in a seated posture is needed for clinical research and practice. Its applications include wheelchair seating evaluations and ergonomic assessments of seated workers.

Entities:  

Mesh:

Year:  2003        PMID: 12763443     DOI: 10.1016/s0268-0033(03)00049-4

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  7 in total

1.  Development of valid and reliable measures of postural stability.

Authors:  Stephen Sprigle; Christine Maurer; Mark Holowka
Journal:  J Spinal Cord Med       Date:  2007       Impact factor: 1.985

Review 2.  Position-sensing technologies for movement analysis in stroke rehabilitation.

Authors:  H Zheng; N D Black; N D Harris
Journal:  Med Biol Eng Comput       Date:  2005-07       Impact factor: 2.602

3.  Posterior pelvic tilt is a risk factor for rectal prolapse: a propensity score matching analysis.

Authors:  C Cantiani; D Sgamma; E Grossi; D Saccá; O R Meli; Q Lai; F Gaj
Journal:  Tech Coloproctol       Date:  2020-03-13       Impact factor: 3.781

4.  Test-re-test reliability and inter-rater reliability of a digital pelvic inclinometer in young, healthy males and females.

Authors:  Chris Beardsley; Tim Egerton; Brendon Skinner
Journal:  PeerJ       Date:  2016-03-31       Impact factor: 2.984

5.  The modified Thomas test is not a valid measure of hip extension unless pelvic tilt is controlled.

Authors:  Andrew D Vigotsky; Gregory J Lehman; Chris Beardsley; Bret Contreras; Bryan Chung; Erin H Feser
Journal:  PeerJ       Date:  2016-08-11       Impact factor: 2.984

6.  Effects of anterior weight-shifting methods on sitting balance in wheelchair-dependent patients with spinal cord injury.

Authors:  Han Seong Choe; Dong-Ki Min; Jeoungah Ahn
Journal:  J Phys Ther Sci       Date:  2018-03-02

7.  A Sensor-Based Screening Tool for Identifying High Pelvic Mobility in Patients Due to Undergo Total Hip Arthroplasty.

Authors:  Xueyang Wang; Arham Qureshi; Abhinav Vepa; Usama Rahman; Arnab Palit; Mark A Williams; Richard King; Mark T Elliott
Journal:  Sensors (Basel)       Date:  2020-10-30       Impact factor: 3.576

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.