OBJECTIVE: This study sought to evaluate whether surgical clips affect tissue conductivity and thereby alter the induction of radiofrequency ablation lesions and to determine whether therapy is safe after previous placement of clips in the liver.
MATERIALS AND METHODS: An ex vivo porcine hepatic model was used. Three clips were placed around a radiofrequency electrode at 10, 20, and 30 mm from the point of insertion. Clips were arranged in a plane either perpendicular or parallel to the electrode track. After placement of the liver specimen on a grounding pad, radiofrequency energy was applied in a standardized manner for 5 min. Lesion growth and morphology were documented for each minute.
RESULTS: Radiofrequency lesions appeared circular and homogeneous after 5 min. Lesion diameter perpendicular to the radiofrequency electrode averaged 30 mm. However, lesion formation was irregular during the early phase of the radiofrequency ablation. The lesion extended irregularly toward the 1-cm clip after 60 sec of ablation. During the second minute, a distinct lesion was observed around the clip 1 cm from the electrode; the primary lesion had not yet reached the clip. During the final 3 min, the primary lesion reached the 1-cm clip and ultimately incorporated the satellite lesion. No lesions were detected surrounding the more distant clips.
CONCLUSION: Our data suggest that with the parameters applied in our study, radiofrequency ablation can be safely performed in patients with implanted clips. No aberrant conduction is observed around surgical clips that are located 20 mm and further from the radiofrequency electrode.

Entities:  

Mesh:

Year:  2003        PMID: 12760918     DOI: 10.2214/ajr.180.6.1801557

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  5 in total

Review 1.  Radiofrequency thermal ablation of liver tumors.

Authors:  Elisabetta Buscarini; Agostino Savoia; Gianfranco Brambilla; Fernanda Menozzi; Luigi Reduzzi; Deike Strobel; Johannes Hänsler; Luigi Buscarini; Luigi Gaiti; Alessandro Zambelli
Journal:  Eur Radiol       Date:  2005-03-08       Impact factor: 5.315

2.  Gel Phantom Models for Radiofrequency and Microwave Ablation of the Liver.

Authors:  Willa J Chen; Qi Wang; Charles Y Kim
Journal:  Dig Dis Interv       Date:  2020-09-28

3.  Safety And Necessity Of Thermal Esophageal Probes During Radiofrequency Ablation For The Treatment Of Atrial Fibrillation.

Authors:  Antonio Fasano; Prof Emeritus; Luca Anfuso; Stefano Bozzi; Claudio PandoziProf
Journal:  J Atr Fibrillation       Date:  2016-06-30

4.  Percutaneous coil placement prior to radiofrequency ablation of poorly visible hepatic tumors.

Authors:  Andreas Adam; Adam Hatzidakis; Mohamad Hamady; Tarun Sabharwal; Afshin Gangi
Journal:  Eur Radiol       Date:  2004-03-20       Impact factor: 5.315

5.  Cryoablation of lung malignancies recurring close to surgical clips following surgery: Report of three cases.

Authors:  Rosario F Grasso; Giacomo Luppi; Roberto L Cazzato; Riccardo Del Vescovo; Francesco Giurazza; Simona Mercurio; Eliodoro Faiella; Bruno Beomonte Zobel
Journal:  Indian J Radiol Imaging       Date:  2015 Jan-Mar
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.