Literature DB >> 1276024

Towards blood-gas autoanalysis an evaluation of the radiometer ABL.

B J Selman, A R Tait.   

Abstract

The performance of the Radiometer ABL 1 blood-gas analyser has been evaluated using tonometered blood samples of known PO2 and PCO2. The pH electrode was compared with two other pH electrode systems and the haemoglobin estimations were compared with measurements made by the standard cyanmethaemoglobin method. The accuracy of the machine compared well with standard blood-gas equipment, but the haemoglobin values had a small error because of a zero offset in the machine. The analyzer is simple to operate and has low running costs.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 1276024     DOI: 10.1093/bja/48.5.487

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  3 in total

1.  Variability of arterial PO2 in hospitalized patients: implications for clinical practice and research.

Authors:  P Pompei; M E Charlson; T K King
Journal:  J Gen Intern Med       Date:  1987 Mar-Apr       Impact factor: 5.128

2.  The metabolic state of muscle in the isolated perfused rat hemicorpus in relation to rates of protein synthesis.

Authors:  V R Preedy; V M Pain; P J Garlick
Journal:  Biochem J       Date:  1984-03-01       Impact factor: 3.857

3.  Oxygen tension measurement using an automatic blood gas analyser.

Authors:  J Becket; C Orchard; M K Chakrabarti; G M Hall; I D Gillies; P J Bourdillon
Journal:  J R Soc Med       Date:  1981-08       Impact factor: 18.000