To what extent can digital images obtained with a non-isocentric C-arm be used for brachytherapy treatment planning in gynaecology.

Jeno Pálvölgyi1.   

Abstract

BACKGROUND AND
PURPOSE: The non-isocentric C-arm X-ray fluoroscopy unit is not typically used as a brachytherapy localizer. The main aim of the study is to examine the extent to which the potential of a mobile C-arm unit as a brachytherapy localizer for gynaecological treatments with Fletcher-Suit applicator can be used. PATIENTS AND METHODS: With a C-arm, located in the brachytherapy treatment room in situ reconstruction images are made, eliminating the transportation of the patient to the X-ray room. The brachytherapy treatment planning is based on the frame grabbed images using variable angle reconstruction.
RESULTS: The method of making a pair of reconstruction images with the C-arm is described. The accuracy of the reconstruction has been examined with phantom studies and therapy plans of Fletcher-Suit applications of different variable angles.

Entities:  

Mesh:

Year:  2003        PMID: 12758246     DOI: 10.1016/s0167-8140(03)00027-6

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  3 in total

1.  Dummy source digitization algorithm for reconstruction of flexible brachytherapy catheters with biplane images.

Authors:  Jenö Pálvölgyi
Journal:  J Contemp Brachytherapy       Date:  2014-02-18

2.  Modified Heyman packing performed with small number of Simon-Norman catheters.

Authors:  Jenö Pálvölgyi; Kofi Agyemang-Prempeh
Journal:  J Contemp Brachytherapy       Date:  2010-04-01

3.  The consistency of Fletcher-Suit applicator geometry and of the rectal probe's position in high dose rate brachytherapy treatment fraction of cervix carcinoma.

Authors:  Jenő Pálvölgyi
Journal:  J Contemp Brachytherapy       Date:  2009-10-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.