| Literature DB >> 12757949 |
Gertie C A M Bokken1, Ronald J Corbee, Frans van Knapen, Aldert A Bergwerff.
Abstract
A surface plasmon resonance biosensor (Biacore) was used to detect Salmonella through antibodies reacting with Salmonella group A, B, D and E (Kauffmann-White typing). In the assay designed, anti-Salmonella antibodies immobilized to the biosensor surface were allowed to bind injected bacteria followed by a pulse with soluble anti-Salmonella immunoglobulins to intensify the signal. No significant interference was found for (mixtures of) 30 non-Salmonella serovars at 10(9) CFU ml(-1). A total of 53 Salmonella serovars were successfully detected at 1 x 10(7) CFU ml(-1), except those of groups C, G, L and P, as expected. The cut-off point was determined with an equicellular mixture of Salmonella enteritidis and Salmonella typhimurium at a final amount of 1.7 x 10(3) CFU per test portion. Although further work is needed to cover the detection of all relevant Salmonella serovars in food-producing animals and food products, this work demonstrates the merits of this alternative biosensor approach in terms of automation, sensitivity, specificity, simple handling and limited hands-on time.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12757949 DOI: 10.1016/S0378-1097(03)00250-7
Source DB: PubMed Journal: FEMS Microbiol Lett ISSN: 0378-1097 Impact factor: 2.742