Evaluation of hematological values obtained with reference automated hematology analyzers of six manufacturers.

Takayuki Takubo1, Noriyuki Tatsumi, Naotake Satoh, Kazuhiko Matsuno, Keiji Fujimoto, Masahiko Soga, Yoshinori Yamagami, Shunichi Akiba, Tohru Sudoh, Makoto Miyazaki.   

Abstract

We evaluated assays of the same fresh blood samples with six different types of reference automated hematology analyzers developed by the following manufacturers: Beckman Coulter, Sysmex, Bayer, Abbott, Nihon Kohden and Horiba. Fresh whole blood samples treated with dipotassium ethylenediaminetetraacetic acid (EDTA K2) were collected from three healthy adult volunteers. The complete blood counts (CBC) including red blood cell count (RBC), hemoglobin (Hgb), hematocrit (Hct), mean corpuscular volume (MCV), white blood cell count (WBC), platelet count (Plt), reticulocyte percentage (Ret) and leukocyte differential counts including % neutrophils (Neu), % lymphocytes (Lym) and % monocytes (Mon) were surveyed with a reference automated hematology analyzer from each manufacturer. The process from sampling to analysis was performed according to procedures in hospital clinical laboratories. RBC, Hgb, Hct and MCV exhibited allowable differences within 5% of mean value among all instruments. Large differences greater than 10% of mean value in WBC, Neu and Lym between Horiba and other manufacturers, and in Plt between Nihon Kohden and other manufacturers, were observed. Ret and Mon exhibited large differences over 10% of mean value among almost all of the instruments tested. This survey suggests that all parameters exhibiting differences greater than 10% of mean value among instruments should be improved for clinical use to ensure good external quality control in blood cell counting and leukocyte differential counting using automated instruments.

Entities:  

Mesh:

Year:  2002        PMID: 12755270

Source DB:  PubMed          Journal:  Southeast Asian J Trop Med Public Health        ISSN: 0125-1562            Impact factor:   0.267


  4 in total

1.  Reference intervals of red blood cell parameters and platelet count for healthy adults in Japan.

Authors:  Akiyoshi Takami; Shinichiro Watanabe; Yoshikazu Yamamoto; Hiroshi Kondo; Yukiharu Bamba; Masahiko Ohata; Seiji Mishima; Hiroshi Kubota; Akihiko Nishiura; Reiko Miura; Miyuki Iwagami; Katsumi Higashi; Yutaka Yatomi; Kaoru Tohyama
Journal:  Int J Hematol       Date:  2021-06-02       Impact factor: 2.490

2.  Evaluation of mouse red blood cell and platelet counting with an automated hematology analyzer.

Authors:  Teruko Fukuda; Eri Asou; Kimiko Nogi; Kazuo Goto
Journal:  J Vet Med Sci       Date:  2017-08-27       Impact factor: 1.267

3.  Effect of Altitude on Hemoglobin and Red Blood Cell Indices in Adults in Different Regions of Saudi Arabia.

Authors:  Husain Y Alkhaldy; Zuhier A Awan; Ahmed A Abouzaid; Hossameldin M Elbahaey; Saeed M Al Amoudi; Shehata F Shehata; Muhammad Saboor
Journal:  Int J Gen Med       Date:  2022-03-31

4.  Comment on "Inflawell, neutrophil-to-lymphocyte ratio and COVID-19".

Authors:  Rujittika Mungmunpuntipantip; Viroj Wiwanitkit
Journal:  Inflammopharmacology       Date:  2022-03-22       Impact factor: 5.093

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.