Literature DB >> 12748352

Reducing vascular access morbidity: a comparative trial of two vascular access monitoring strategies.

Charmaine E Lok1, Cynthia Bhola, Ruth Croxford, Robert M A Richardson.   

Abstract

BACKGROUND: Thrombosis is the primary cause of access failure in polytetrafluoroethylene grafts and arteriovenous fistulas. It can lead to significant patient and access morbidity and mortality, and is difficult to prevent medically. Intervention is largely limited to maximizing access patency by detecting culprit lesions early and intervening with angioplasty or surgical revision. The most efficacious monitoring strategy is undetermined.
METHODS: This 3 year prospective study took advantage of a change in monitoring strategy used in a large dialysis centre to compare the efficacy of two methods used to monitor grafts and fistulas in order to prevent access thrombosis. Accesses were monitored using Duplex ultrasonography in year 1, while the saline ultrasound dilution technique (Transonic) became the primary monitoring strategy in year 3 (year 2 was a transition year). Risk factors for thrombosis were determined using multivariate survival analysis, and the performance of Duplex ultrasonography and Transonic monitoring was assessed.
RESULTS: A total of 303 656 access days at risk were assessed, with 344, 385 and 425 accesses in years 1, 2 and 3, respectively. The total thrombosis rate was 1.01/1000 access days in year 1 compared with 0.66/1000 access days in year 3. This was accomplished despite a reduction in procedure rates of 55% for angiograms, 13% for angioplasties and 31% for thrombolysis.
CONCLUSION: Low flow rates detected using Transonic monitoring were associated with increased thrombosis, while stenosis detected using Duplex ultrasonography was not a strong predictor of incipient thrombosis; however, these different access characteristics were compared using monitoring techniques that may be ideal in different clinical situations.

Entities:  

Mesh:

Year:  2003        PMID: 12748352     DOI: 10.1093/ndt/gfg122

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  9 in total

Review 1.  Combined functional and anatomical diagnostic endpoints for assessing arteriovenous fistula dysfunction.

Authors:  Ehsan Rajabi-Jaghargh; Rupak K Banerjee
Journal:  World J Nephrol       Date:  2015-02-06

2.  Optimizing function and treatment of hemodialysis grafts and fistulae.

Authors:  Thomas M Vesely
Journal:  Semin Intervent Radiol       Date:  2004-06       Impact factor: 1.513

Review 3.  Complications of the Arteriovenous Fistula: A Systematic Review.

Authors:  Ahmed A Al-Jaishi; Aiden R Liu; Charmaine E Lok; Joyce C Zhang; Louise M Moist
Journal:  J Am Soc Nephrol       Date:  2016-12-28       Impact factor: 10.121

4.  Does regular surveillance improve the long-term survival of arteriovenous fistulas?

Authors:  Ashwin Shetty; William L Whittier
Journal:  Int J Nephrol       Date:  2011-12-01

5.  Effect of online haemodialysis vascular access flow evaluation and pre-emptive intervention on the frequency of access thrombosis.

Authors:  Edwin Wijnen; Frank M van der Sande; Jan H M Tordoir; Jeroen P Kooman; Karel M L Leunissen
Journal:  NDT Plus       Date:  2008-08-22

6.  Hemodialysis Infection Prevention Protocols Ontario-Shower Technique (HIPPO-ST): A Pilot Randomized Trial.

Authors:  S Daisy Kosa; Amiram Gafni; Andrew A House; JulieAnn Lawrence; Louise Moist; Bharat Nathoo; Paul Tam; Alicia Sarabia; Lehana Thabane; George Wu; Charmaine E Lok
Journal:  Kidney Int Rep       Date:  2016-11-08

7.  Association of vascular access flow with short-term and long-term mortality in chronic haemodialysis patients: a retrospective cohort study.

Authors:  Chung-Kuan Wu; Chia-Lin Wu; Chia-Hsun Lin; Jyh-Gang Leu; Chew-Teng Kor; Der-Cherng Tarng
Journal:  BMJ Open       Date:  2017-09-24       Impact factor: 2.692

8.  Weight-Based Assessment of Access Flow Threshold to Predict Arteriovenous Fistula Functional Patency.

Authors:  Chih-Yu Yang; Bo-Sheng Wu; Yi-Fang Wang; Yan-Hwa Wu Lee; Der-Cherng Tarng
Journal:  Kidney Int Rep       Date:  2021-11-24

Review 9.  Comparison of ultrasound scan blood flow measurement versus other forms of surveillance in the thrombosis rate of hemodialysis access: A systemic review and meta-analysis.

Authors:  Seun Deuk Hwang; Jin Ho Lee; Seoung Woo Lee; Joong Kyung Kim; Moon-Jae Kim; Joon Ho Song
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.