| Literature DB >> 12748107 |
Perry P Ng1, Van V Halbach, Louis P Caragine, Christopher F Dowd, Randall T Higashida.
Abstract
We present a case of Guglielmi detachable coil extraction into a microcatheter after coil detachment during embolization of a superior hypophyseal artery aneurysm; extraction was a result of the suction generated during pusher-wire withdrawal. Experimental simulations using many coil and microcatheter combinations were used to identify factors that contribute to the likelihood of this phenomenon.Entities:
Mesh:
Year: 2003 PMID: 12748107 PMCID: PMC7975790
Source DB: PubMed Journal: AJNR Am J Neuroradiol ISSN: 0195-6108 Impact factor: 3.825