On the practice of the clinical implementation of enhanced dynamic wedges.

Phil W Koken1, Stan Heukelom, Johan P Cuijpers.   

Abstract

Practical aspects of the clinical implementation of enhanced dynamic wedges (EDW) replacing manual wedges are presented and discussed extensively. A comparison between measured and calculated data is also presented. Relative dose distributions and wedge factors were calculated with a commercially available treatment planning system and measured in a water-phantom and with an ionization chamber. Wedge factor calculations and measurements were also compared with an independent method of wedge factor calculations available from the literature. Aspects of the clinical implementation, such as safety and quality assurance, were evaluated. Measurements and calculations agreed very well and were slightly better than results of previous studies. Profiles and percentage depth doses (PDDs) agreed within 1% to 1.5% and within 0.5%, respectively. Measured and calculated wedge factors ratios agreed within 0.5% to 1%. Calculated and measured EDW dose distributions showed excellent agreement, both relative and absolute. However, for safe and practical use, specific aspects need to be taken into consideration. Once the treatment planning system is commissioned properly, the clinical implementation of EDW is rather straightforward.

Entities:  

Mesh:

Year:  2003        PMID: 12747613     DOI: 10.1016/S0958-3947(02)00140-1

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database.

Authors:  Christopher F Njeh
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

2.  Use of a multileaf collimator to increase the field width achievable with a dynamic wedge.

Authors:  Ivan A Brezovich; Richard A Popple; Jun Duan
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

3.  Performance evaluation and quality assurance of Varian enhanced dynamic wedges.

Authors:  Parham Alaei; Patrick D Higgins
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.