Literature DB >> 12728033

Cyclic loading of olecranon fracture fixation constructs.

Douglas T Hutchinson1, Daniel S Horwitz, Gregory Ha, Cameron W Thomas, Kent N Bachus.   

Abstract

BACKGROUND: Despite the good results that are usually reported after fixation at the sites of olecranon fractures and osteotomies, problems such as loss of fixation, nonunion, and the need for revision surgery are still encountered. Various types of fixation have been recommended, but few have been evaluated with use of clinically relevant cyclic load testing at appropriate levels of stress. The purpose of the present study was to test multiple olecranon fixation techniques under physiologic cyclic loads.
METHODS: We studied ten cadaveric elbows with use of cyclic loading that simulated (1) active range of motion and (2) pushing up from a chair. Each specimen underwent fixation of a simulated 50% transverse olecranon fracture with use of intramedullary and cortically fixed tension band constructs (in randomized order) followed by fixation with a 7.3-mm-diameter cancellous screw with and without a tension band. Displacement transducers were placed posteriorly on the tension side and anteriorly near the articular surface.
RESULTS: Both configurations involving the 7.3-mm-diameter cancellous screw provided the most stable fixation-nearly five times better than that provided by the Kirschner-wire techniques. Use of the tension band in conjunction with the intramedullary screw improved the stability of fixation. In none of the constructs did the AO tension band result in compression across the osteotomy gap. CONCLUSIONS AND CLINICAL RELEVANCE: The use of a 7.3-mm screw in conjunction with a tension band provided better fixation of simulated displaced transverse fractures than did the use of Kirschner wires in conjunction with a tension band or the use of a screw only. The AO principle of converting posterior tensile forces to articular compressive forces was not demonstrated in this study. We therefore question the validity of the tension band concept in olecranon fracture fixation and recommend passive rather than active range of motion in the immediate postoperative period to limit fracture distraction.

Entities:  

Mesh:

Year:  2003        PMID: 12728033     DOI: 10.2106/00004623-200305000-00010

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  24 in total

1.  Material properties of common suture materials in orthopaedic surgery.

Authors:  S Najibi; R Banglmeier; Jm Matta; M Tannast
Journal:  Iowa Orthop J       Date:  2010

Review 2.  Fractures in brief: Olecranon fractures.

Authors:  Laura Wiegand; Joseph Bernstein; Jaimo Ahn
Journal:  Clin Orthop Relat Res       Date:  2012-12       Impact factor: 4.176

3.  Repair of comminuted (Mayo type IIB) olecranon fracture using Ethi-bond 5 sutures without metallic implants: A novel technique.

Authors:  Sheng Xu; Heng An Lin; Merng Koon Wong
Journal:  J Orthop       Date:  2019-03-12

4.  Does K-wire position in tension band wiring of olecranon fractures affect its complications and removal of metal rate?

Authors:  K W Chan; K J Donnelly
Journal:  J Orthop       Date:  2014-06-17

5.  CORR Insights®: Long-term outcome of displaced, transverse, noncomminuted olecranon fractures.

Authors:  Boris A Zelle
Journal:  Clin Orthop Relat Res       Date:  2014-03-11       Impact factor: 4.176

6.  [Olecranon osteotomy].

Authors:  R Meier; F Gohlke
Journal:  Orthopade       Date:  2013-05       Impact factor: 1.087

7.  Outcome after olecranon fracture repair: Does construct type matter?

Authors:  Edward M DelSole; Christian A Pean; Nirmal C Tejwani; Kenneth A Egol
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-11-16

8.  Outcome of isolated olecranon fractures in skeletally immature patients: comparison of open reduction and tension band wiring fixation versus closed reduction and percutaneous screw fixation.

Authors:  Marco Corradin; Lorenza Marengo; Antonio Andreacchio; Matteo Paonessa; Vanessa Giacometti; Antoine Samba; Marie Rousset; Federico Canavese
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-04-30

9.  Design and application of nickel-titanium olecranon memory connector in treatment of olecranon fractures: a prospective randomized controlled trial.

Authors:  Xiao Chen; Peng Liu; Xiaofei Zhu; Liehu Cao; Chuncai Zhang; Jiacan Su
Journal:  Int Orthop       Date:  2013-04-18       Impact factor: 3.075

10.  Biomechanical evaluation of fixation of comminuted olecranon fractures: one-third tubular versus locking compression plating.

Authors:  Geert A Buijze; Leendert Blankevoort; Gabriëlle J M Tuijthof; Inger N Sierevelt; Peter Kloen
Journal:  Arch Orthop Trauma Surg       Date:  2009-10-13       Impact factor: 3.067