| Literature DB >> 12720358 |
Tanner M Schaub1, Christopher L Hendrickson, Kuangnan Qian, John P Quinn, Alan G Marshall.
Abstract
We report the first field desorption ionization broadband high-resolution (m/Deltam(50%) approximately 65 000) mass spectra. We have interfaced a field ionization/field desorption source to a home-built 9.4-T FT-ICR mass spectrometer. The instrumental configuration employs convenient sample introduction (in-source liquid injection) and external ion accumulation. We demonstrate the utility of this configuration by generating high-resolution positive-ion mass spectra of C(60) and a midboiling crude oil distillate. The latter contains species not accessible by common soft-ionization methods, for example, low-voltage electron ionization, electrospray ionization, and matrix-assisted laser desorption/ionization. The present work demonstrates significant advantages of FI/FD FT-ICR MS for analysis of nonpolar molecules in complex mixtures.Entities:
Year: 2003 PMID: 12720358 DOI: 10.1021/ac020627v
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986