Randomized comparison of the coil-design Crossflex and the tubular NIR stent.

Leif Thuesen1, Henning R Andersen, Lars R Krusell, Hans E Bøtker, Erik Jørgensen, Henning Kelbaek, Steen D Kristensen.   

Abstract

The purpose of this study was to compare the angiographic outcome of implantation of the coil-design Crossflex stent with the tubular NIR stent for treatment of coronary artery stenoses. Two hundred twenty-three patients with one genuine coronary artery lesion were randomized to implantation with a 15 mm Crossflex stent (n = 112) or a 16 mm NIR stent (n = 111). The patients had angiographic follow-up after 6 months. Primary endpoint was minimal luminal diameter (MLD) after 6 months. There was a similar clinical outcome in the two groups. At 6-month follow-up, the MLD was significantly lower in the Crossflex group (1.94 +/- 0.79 mm) than in the NIR group (2.37 +/- 0.84 mm; P < 0.001). Early gain was the same in the two groups. Late loss and percent diameter stenoses were significantly higher in the Crossflex group. The binary restenoses rate was 26% and 17% in the Crossflex and the NIR groups, respectively (P = NS). The coil-design Crossflex stent was found to be inferior to the tubular NIR stent concerning late loss and MLD at 6-month follow-up. Copyright 2003 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2003        PMID: 12720233     DOI: 10.1002/ccd.10499

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

Review 1.  The development of carotid stent material.

Authors:  Dongsheng He; Wenhua Liu; Tao Zhang
Journal:  Interv Neurol       Date:  2015-03