| Literature DB >> 12713918 |
Marc Delvaux1, Sophie Demoustier-Champagne.
Abstract
Arrays of nanoscopic gold tubes were prepared by electroless deposition of the metal within the pores of polycarbonate particle track-etched membranes (PTM). Glucose oxidase (Gox) was immobilised onto preformed self-assembled monolayers (SAMs) (mercaptoethylamine or mercaptopropionic acid) on electroless gold via cross-linking with glutaraldehyde or covalent attachment by carbodiimide coupling. The effectiveness of the different steps in the Gox immobilisation procedure was assessed by contact angle measurements, cyclic voltammetry and X-ray photoelectron spectroscopy. The enzyme loading was estimated by radioactivity measurements. The sensitivity to beta-glucose of these different biosensors has been evaluated. Glucose responses as large as 400 nA mM(-1) cm(-2) have been obtained. To our knowledge, this sensitivity value is amongst the highest values reported in the literature for comparable biosensor systems.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12713918 DOI: 10.1016/s0956-5663(02)00209-9
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618