The self-inflating bulb as an esophageal detector device in children weighing more than twenty kilograms: a comparison of two techniques.

Ghazala Q Sharieff1, Alexander Rodarte, Niall Wilton, Patricia D Silva, Dianne Bleyle.   

Abstract

STUDY
OBJECTIVES: We confirm the ability of the self-inflating bulb to indicate endotracheal tube position in children and determine which method of bulb compression is more accurate.
METHODS: This single-blind, prospective, single cohort, repeated measures comparison of the on-deflate and the off-deflate methods of bulb compression was conducted in the operating room of a children's hospital. Seventy-five patients weighing more than 20 kg were enrolled. All patients had an endotracheal tube placed in both the trachea and the esophagus, 5 mL/kg of air was insufflated into the stomach, and 4 measurements were obtained on each patient. The bulb was either applied to the endotracheal tube and then compressed (on-deflate method), or compressed before its attachment to the endotracheal tube (off-deflate method). The order in which the methods were utilized was determined by a computer-generated permuted block randomization scheme. The blinded assessor told the anesthesiologist when to start and stop the clock after each intervention. Five seconds was used as the cut-off time for which the clock was stopped. If the bulb reexpanded within 5 seconds, then the tube was considered to be in the trachea; if it did not reexpand, then this was considered to be an esophageal intubation.
RESULTS: The mean patient age was 11 years. The on-deflate method had a sensitivity of 99% and a specificity of 92% for detecting tracheal intubations. The off-deflate method had a sensitivity of 99% and a specificity of 100% (95% confidence interval [CI] for the difference between methods: sensitivity -6 to 6, specificity -14 to 0.4). The inaccuracy rate for the off-deflate method was 1%, whereas the inaccuracy rate for the on-deflate method was 5% (Delta4; 95% CI 0.2 to 9).
CONCLUSION: The self-inflating bulb is a reliable method of detecting endotracheal tube position in children. Furthermore, the off-deflate method may be more reliable and accurate than the on-deflate method when used in children weighing more than 20 kg.

Entities:  

Mesh:

Year:  2003        PMID: 12712028     DOI: 10.1067/mem.2003.185

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  3 in total

1.  Part 10: Pediatric basic and advanced life support: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations.

Authors:  Monica E Kleinman; Allan R de Caen; Leon Chameides; Dianne L Atkins; Robert A Berg; Marc D Berg; Farhan Bhanji; Dominique Biarent; Robert Bingham; Ashraf H Coovadia; Mary Fran Hazinski; Robert W Hickey; Vinay M Nadkarni; Amelia G Reis; Antonio Rodriguez-Nunez; James Tibballs; Arno L Zaritsky; David Zideman
Journal:  Circulation       Date:  2010-10-19       Impact factor: 29.690

2.  Pediatric basic and advanced life support: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations.

Authors:  Monica E Kleinman; Allan R de Caen; Leon Chameides; Dianne L Atkins; Robert A Berg; Marc D Berg; Farhan Bhanji; Dominique Biarent; Robert Bingham; Ashraf H Coovadia; Mary Fran Hazinski; Robert W Hickey; Vinay M Nadkarni; Amelia G Reis; Antonio Rodriguez-Nunez; James Tibballs; Arno L Zaritsky; David Zideman
Journal:  Pediatrics       Date:  2010-10-18       Impact factor: 7.124

3.  Canadian Airway Focus Group updated consensus-based recommendations for management of the difficult airway: part 1. Difficult airway management encountered in an unconscious patient.

Authors:  J Adam Law; Laura V Duggan; Mathieu Asselin; Paul Baker; Edward Crosby; Andrew Downey; Orlando R Hung; Philip M Jones; François Lemay; Rudiger Noppens; Matteo Parotto; Roanne Preston; Nick Sowers; Kathryn Sparrow; Timothy P Turkstra; David T Wong; George Kovacs
Journal:  Can J Anaesth       Date:  2021-06-18       Impact factor: 5.063

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.