Abstract

The purpose of this work was to determine the within-session variability and between-session repeatability of the Canon Laser Blood Flowmeter (CLBF), Model 100, an instrument that permits the noninvasive measurement of retinal blood flow. The CLBF calculates flow in microl/min based on the Poiseuille principle. One eye of 20 normal subjects (mean age 36.5; SD 9.7 years) was randomly chosen. A minimum of five measurements was acquired of a temporal arteriole approximately 1 disc diameter from the optic nerve. Measurements were repeated within a 1-month period. Blood pressure and intraocular pressure were measured. The coefficient of variation (COV) and the coefficient of repeatability (COR) were calculated for each individual. The individual COVs for diameter, velocity, and flow ranged from 0.5 to 6.5% (median 2.0%), 4.8 to 39.7% (median 19.9%), and 4.8 to 37.3% (median 19.3%), respectively. The group mean CORs for diameter, velocity, and flow were 5.2 microm (relative to a mean effect of 104.6 microm), 8.8 mm/s (relative to a mean effect of 33.9 mm/s), and 2.6 microl/min (relative to a mean effect of 8.8 microl/min), respectively. The CLBF gave consistent and repeatable measurements of blood flow within retinal arterioles in normal subjects. Given the range of individual variation in the velocity measurement, and thus flow, confidence limits for retinal hemodynamics need to be determined on an individual basis.

Mesh:

Year:  2003        PMID: 12711255     DOI: 10.1016/s0026-2862(03)00007-4

Source DB:  PubMed          Journal:  Microvasc Res        ISSN: 0026-2862            Impact factor:   3.514


  16 in total

Review 1.  Retinal vascular image analysis as a potential screening tool for cerebrovascular disease: a rationale based on homology between cerebral and retinal microvasculatures.

Authors:  Niall Patton; Tariq Aslam; Thomas Macgillivray; Alison Pattie; Ian J Deary; Baljean Dhillon
Journal:  J Anat       Date:  2005-04       Impact factor: 2.610

2.  Reliability of vessel diameter measurements with a retinal oximeter.

Authors:  Renata Blondal; Margret Kara Sturludottir; Sveinn Hakon Hardarson; Gisli Hreinn Halldorsson; Einar Stefánsson
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-04-16       Impact factor: 3.117

3.  Cardiac-Gated En Face Doppler Measurement of Retinal Blood Flow Using Swept-Source Optical Coherence Tomography at 100,000 Axial Scans per Second.

Authors:  ByungKun Lee; WooJhon Choi; Jonathan J Liu; Chen D Lu; Joel S Schuman; Gadi Wollstein; Jay S Duker; Nadia K Waheed; James G Fujimoto
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-04       Impact factor: 4.799

4.  Dynamic Alterations in Blood Flow in Glaucoma Measured with Laser Speckle Contrast Imaging.

Authors:  Alfred Vinnett; Jayanth Kandukuri; Christopher Le; Kyoung-A Cho; Avigyan Sinha; Samuel Asanad; Ginger Thompson; Victoria Chen; Abhishek Rege; Osamah J Saeedi
Journal:  Ophthalmol Glaucoma       Date:  2021-10-18

5.  Reduction of retinal blood flow in high myopia.

Authors:  Noriaki Shimada; Kyoko Ohno-Matsui; Seiyo Harino; Takeshi Yoshida; Kenjiro Yasuzumi; Ariko Kojima; Kanako Kobayashi; Soh Futagami; Takashi Tokoro; Manabu Mochizuki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-01-14       Impact factor: 3.117

6.  Relationship between retinal blood flow and arterial oxygen.

Authors:  Richard W Cheng; Firdaus Yusof; Edmund Tsui; Monica Jong; James Duffin; John G Flanagan; Joseph A Fisher; Chris Hudson
Journal:  J Physiol       Date:  2015-12-30       Impact factor: 5.182

7.  Noninvasive Assessment of Retinal Blood Flow Using a Novel Handheld Laser Speckle Contrast Imager.

Authors:  Abhishek Rege; Samantha I Cunningham; Yusi Liu; Karan Raje; Sachin Kalarn; M Jason Brooke; Lisa Schocket; Sunni Scott; Asifa Shafi; Luis Toledo; Osamah J Saeedi
Journal:  Transl Vis Sci Technol       Date:  2018-11-14       Impact factor: 3.283

8.  Repeatability of the evaluation of systemic microvascular endothelial function using laser doppler perfusion monitoring: clinical and statistical implications.

Authors:  Eduardo Tibiriçá; Alessandra S M Matheus; Bruno Nunes; Sandro Sperandei; Marilia B Gomes
Journal:  Clinics (Sao Paulo)       Date:  2011       Impact factor: 2.365

9.  Measurement of Retinal Blood Flow Using Fluorescently Labeled Red Blood Cells.

Authors:  Tess E Kornfield; Eric A Newman
Journal:  eNeuro       Date:  2015 Mar-Apr

10.  En Face Doppler Optical Coherence Tomography Measurement of Total Retinal Blood Flow in Diabetic Retinopathy and Diabetic Macular Edema.

Authors:  ByungKun Lee; Eduardo A Novais; Nadia K Waheed; Mehreen Adhi; Talisa E de Carlo; Emily D Cole; Eric M Moult; WooJhon Choi; Mark Lane; Caroline R Baumal; Jay S Duker; James G Fujimoto
Journal:  JAMA Ophthalmol       Date:  2017-03-01       Impact factor: 7.389