Fibrin sealing improves stability of corneal prostheses during vitreoretinal procedures.

Constantin E Uhlig1, Heinrich Gerding.   

Abstract

PURPOSE: The aim of this study was to test the effectiveness of using a fibrin sealant as an aid to stabilize temporarily sutured keratoprostheses.
METHODS: Ex vivo the corneas were removed from six porcine eyes, and an Eckardt prosthesis was sutured and additionally fixed with fibrin sealant. The outflow resistance was recorded from eyes with and without fibrin-sealed keratoprostheses at different levels of intraocular pressure. The method of sealing the prosthesis was applied in the clinic and documented in four patients during intravitreal surgery.
RESULTS: The flow volume of the irrigating system was measured ex vivo. Control trials were performed to investigate the effects of graduated increases in hydrostatic pressure on the system. Leakage areas were calculated, and the stability of the system was monitored during the surgical procedure. Slit-lamp biomicroscopy, funduscopy, and visual outcome were documented in the patients. Ex vivo the differences in the leakage areas between the fibrin-sealed and the unsealed conditions were statistically significant up to 51.45 mmHg (P = 0.01). In clinical applications, the fibrin sealant stabilized the keratoprosthesis, and no significant leakage or system instabilities occurred.
CONCLUSIONS: The results confirm that sealing increases the stability of the keratoprosthesis and may enable greater surgical control.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12707601     DOI: 10.1097/00006982-200304000-00012

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  4 in total

1.  Fibrin glue in ophthalmology.

Authors:  Niranjan Kumar; Khalid Al Sabti
Journal:  Indian J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 1.848

2.  Can we overcome the challenges of sutures in lamellar keratoplasty?

Authors:  Anita Panda; Sandeep Kumar
Journal:  Indian J Ophthalmol       Date:  2011 Jul-Aug       Impact factor: 1.848

3.  Versatility of use of fibrin glue in wound closure and vitreo-retinal surgery.

Authors:  Gabriela Lopezcarasa-Hernandez; Jose-Francisco Perez-Vazquez; Jose-Luis Guerrero-Naranjo; Maria A Martinez-Castellanos
Journal:  Int J Retina Vitreous       Date:  2021-04-15

Review 4.  Fibrin glue in ophthalmology.

Authors:  Anita Panda; Sandeep Kumar; Abhiyan Kumar; Raseena Bansal; Shibal Bhartiya
Journal:  Indian J Ophthalmol       Date:  2009 Sep-Oct       Impact factor: 1.848

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.