Lutz Bräutigam1, Jens U Nefflen, Gerd Geisslinger.
Abstract
The validation of a liquid chromatography-tandem mass spectrometry (LC-MS-MS) method for the determination of the selective cyclooxygenase-2 inhibitor etoricoxib in human plasma with phenazone as internal standard is described. The plasma samples were extracted by solid-phase extraction using polymer-based cartridges. Chromatography was carried out on a short, narrow bore RP C(18) column (30x2 mm). Detection was achieved by a Sciex API 3000 triple quadrupole mass spectrometer equipped with a turbo ion spray source working in positive ion mode. The respective mass transitions used for quantification of etoricoxib and phenazone were m/z 359.2-->280.2 and m/z 189.0-->104.1. The analytical method was validated over the concentration range 0.2-200 ng/ml. The limit of quantification was 0.2 ng/ml. The method is applicable to pharmacokinetic studies in humans.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12705971 DOI: 10.1016/s1570-0232(03)00034-5
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205