Evaluation of Cobas Integra 800 under simulated routine conditions in six laboratories.

Francisco L Redondo1, Pilar Bermudez, Claudio Cocco, Francesca Colella, Maria Stella Graziani, Walter Fiehn, Thomas Hierla, Gisèle Lemoël, AnneMarie Belliard, Dieudonne Manene, Mourad Meziani, Maryann Liebel, Matthew J McQueen, Wolfgang Stockmann.   

Abstract

The new selective access analyser Cobas Integra 800 from Roche Diagnostics was evaluated in an international multicentre study at six sites. Routine simulation experiments showed good performance and full functionality of the instrument and provocation of anomalous situations generated no problems. The new features on Cobas Integra 800, namely clot detection and dispensing control, worked according to specifications. The imprecision of Cobas Integra 800 fulfilled the proposed quality specifications regarding imprecision of analytical systems for clinical chemistry with few exceptions. Claims for linearity, drift, and carry-over were all within the defined specifications, except urea linearity. Interference exists in some cases, as could be expected due to the chemistries applied. Accuracy met the proposed quality specifications, except in some special cases. Method comparisons with Cobas Integra 700 showed good agreement; comparisons with other analysis systems yielded in several cases explicable deviations. Practicability of Cobas Integra 800 met or exceeded the requirements for more than 95% of all attributes rated. The strong points of the new analysis system were reagent handling, long stability of calibration curves, high number of tests on board, compatibility of the sample carrier to other Roche systems, and the sample integrity check for more reliable analytical results. The improvement of the workflow offered by the 5-position rack and STAT handling like on Cobas Integra 800 makes the instrument attractive for further consolidation in the medium-sized laboratory, for dedicated use of special analytes, and/or as back-up in the large routine laboratory.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12705348     DOI: 10.1515/CCLM.2003.058

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Interference testing.

Authors:  Goce Dimeski
Journal:  Clin Biochem Rev       Date:  2008-08

2.  Hyponatremia as a prognostic and predictive factor in metastatic renal cell carcinoma.

Authors:  A N Jeppesen; H K Jensen; F Donskov; N Marcussen; H von der Maase
Journal:  Br J Cancer       Date:  2010-02-09       Impact factor: 7.640

3.  An innovative approach to functionality testing of analysers in the clinical laboratory.

Authors:  Wolfgang Stockmann; Werner Engeldinger; Albert Kunst; Margaret McGovern
Journal:  J Autom Methods Manag Chem       Date:  2008

4.  Increasing Efficiency and Quality by Consolidation of Clinical Chemistry and Immunochemistry Systems with MODULAR ANALYTICS SWA.

Authors:  Paolo Mocarelli; Gary L Horowitz; Pier Mario Gerthoux; Rossana Cecere; Roland Imdahl; Janneke Ruinemans-Koerts; Hilmar Luthe; Silvia Pesudo Calatayud; Marie Luisa Salve; Albert Kunst; Margaret McGovern; Katherine Ng; Wolfgang Stockmann
Journal:  J Autom Methods Manag Chem       Date:  2008
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.