Synthetic oligonucleotides (ODNs) are routinely analyzed using capillary gel electrophoresis (CGE) for size-sieving based separations as well as electrospray mass spectrometry (ESI-MS) for identification. On-line coupling of these methods is therefore desired in order to combine the analytical capabilities provided by both methods. Performance of on-line CGE-ESI-MS systems is influenced by various parameters, and choice of optimal conditions is crucial for successful coupling experiments. In this study, we explore characteristics of the on-line coupled CGE-ESI-MS system for ODN analysis. Effects of CGE buffer concentration, capillary length, separation and orifice voltage on CGE separation and MS detection of a phosphodiester ODN mixture were examined. Attention was paid to the influence of the interface, such as geometry of capillary alignment, sheath liquid flow-rate and sheath liquid composition on performance of the system.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12703907 DOI: 10.1016/s0021-9673(03)00208-5
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759