Volume-guaranteed ventilation.

Cheryl Riley1, Jobeth Pilcher.   

Abstract

Pressure-limited, time-cycled ventilation has been the primary mode of ventilation for neonates for several decades. But the realization that volume rather than pressure causes ventilator-induced lung injury has led to the development of new strategies for ventilation. Volume guarantee is a mode of ventilation that automatically adjusts the inspiratory pressure to achieve a set tidal volume according to changes in lung compliance or resistance or the patient's respiratory drive. Volume-guaranteed ventilation delivers a specific, preset volume of gas, and inspiration ends when it has been delivered. This mode of ventilation requires careful attention to the infant and to ventilator settings.

Entities:  

Mesh:

Year:  2003        PMID: 12696722     DOI: 10.1891/0730-0832.22.2.17

Source DB:  PubMed          Journal:  Neonatal Netw        ISSN: 0730-0832


  2 in total

1.  Volume-targeted modes of modern neonatal ventilators: how stable is the delivered tidal volume?

Authors:  Thomas Jaecklin; Denis R Morel; Peter C Rimensberger
Journal:  Intensive Care Med       Date:  2006-11-22       Impact factor: 17.440

2.  A comparison of synchronized intermittent mandatory ventilation and pressure-regulated volume control ventilation in elderly patients with acute exacerbations of COPD and respiratory failure.

Authors:  Suchi Chang; Jindong Shi; Cuiping Fu; Xu Wu; Shanqun Li
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2016-05-17
  2 in total