Endotoxins in ophthalmic viscosurgical devices.

H Burkhard Dick1, A J Augustin, T Pakula, N Pfeiffer.   

Abstract

PURPOSE: To measure the endotoxin concentration (EC) of 25 commercially available, hyaluronic acid- and hydroxypropylmethylcellulose-based (HPMC) ophthalmic viscosurgical devices (OVDs).
METHODS: The in vitro Limulus amebocyte lysate (LAL) assay, which indicates the presence of endotoxins originating from gram-negative bacteria, was used to determine the EC. The procedure was performed according to the European Pharmacopoeia/USP. EC including duplicate determinations, negative controls, dilution series with control standard endotoxin, dilution series with sample extract and positive sample control.
RESULTS: 16 OVDs (Amvisc, Amvisc Plus, Biolon, Coatel, Healon, Healon GV, Healon, HPMC Ophtal L, Microvisc, Microvisc Plus, Ocucoat, Provisc, Rayvisc, Viscoat, Visco Shield 2%, Visko 1.4%) had an EC under 1.2 endotoxin units/mL, five (Adatocel, HPMC Ophtal H, LA Gel, Viscorneal, Viscorneal Plus) had an EC > or = 1.2 and < or = 24 EU/ml, and four (Biocorneal, Dispasan also named Ophthalin, Dispasan Plus, Visko 1%) had an EC of > 24 EU/ml. DISCUSSION: To avoid viscoelastic-related inflammatory or immunological reactions, the use of pure OVDs is recommended, especially for surgical procedures with an inherent possibility of leaving viscoelastic remnants in the eye (e.g., cataract surgery, visco-canalostomy or penetrating keratoplasty).

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12696637     DOI: 10.1177/112067210301300209

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  3 in total

1.  Determination of a No Observable Effect Level for Endotoxin Following a Single Intravitreal Administration to Cynomolgus Monkeys.

Authors:  Vladimir Bantseev; Paul E Miller; T Michael Nork; Carol A Rasmussen; Aija McKenzie; Brian J Christian; Helen Booler; Evan A Thackaberry
Journal:  J Ocul Pharmacol Ther       Date:  2019-04-09       Impact factor: 2.671

2.  Effects of phacoemulsification with versus without viscoelastic devices on surgical outcomes.

Authors:  İbrahim Taşkın; Lokman Aslan
Journal:  Int Ophthalmol       Date:  2017-05-09       Impact factor: 2.031

3.  Toxic anterior segment syndrome after foldable artiflex iris-fixated phakic intraocular lens implantation.

Authors:  Lucien A M van Philips
Journal:  J Ophthalmol       Date:  2011-06-05       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.