PURPOSE: To describe a case series to determine the incidence of lens opacities after posterior chamber phakic intraocular lens (IOL) implantation (STAAR Surgical, Monrovia, CA) for very high ametropias.
DESIGN: Retrospective, noncomparative, interventional case series. PARTICIPANTS: Fourteen eyes of 170 consecutive eyes with high ametropias in whom lens opacities developed after posterior chamber phakic IOL implant (PCPIOL). INTERVENTION: Posterior chamber phakic intraocular lens implant. MAIN OUTCOME MEASURES: Lens opacity appearance, localization, and clinical course.
RESULTS: Fourteen eyes developed lens opacities 125 +/- 116 days after phakic IOL implant. All eyes had anterior subcapsular opacities, and two eyes also developed nuclear sclerosis. The anterior opacities did not extend posteriorly within the lens, and there were no posterior subcapsular cataracts. Seventy-one percent of opacities were first seen </=3 months, and 86% were seen </=7 months postoperatively. Seventy-nine percent of opacities were seen in the first or second implants of surgeons being trained; 19% of the first 16 cases and 0% of the next 43 cases of one surgeon developed opacities. Mean follow-up after opacity diagnosis was 9.1 +/- 6.8 months. Nine of the 14 (64%) opacities were asymptomatic. Two eyes developed nocturnal glare, with no loss of best-corrected visual acuity (BCVA) and one had loss of BCVA. Two additional eyes with both nuclear sclerotic and anterior subcapsular lens opacities had visual symptoms and/or loss of BCVA and underwent phakic IOL explantation, cataract extraction by phacoemulsification, and PCPIOL implant with good visual outcome. In the entire series, 5 of 170 (2.3%) implantations had symptomatic opacities in which 111 implantations were the first or second case of the implanting surgeon.
CONCLUSIONS: Lens opacities are a potential complication of phakic IOL implantation. Most lens opacities were first seen in the early postoperative period and were most likely due to surgically induced trauma. The anterior subcapsular type was most common and tended not to be rapidly progressive during the follow-up period. The presence of nuclear sclerotic cataract was visually significant and progressive. Long-term follow-up is warranted to evaluate the rate of progression and course of lens opacities after phakic IOL implant surgery.

Entities:  

Mesh:

Year:  2003        PMID: 12689902     DOI: 10.1016/s0161-6420(02)01973-5

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  26 in total

1.  Changes in the crystalline lens resulting from insertion of a phakic IOL (ICL) into the porcine eye.

Authors:  Kunitoshi Fujisawa; Kimiya Shimizu; Shigekazu Uga; Masanobu Suzuki; Koichi Nagano; Yuuki Murakami; Hiroko Goseki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-04-25       Impact factor: 3.117

Review 2.  [Phakic intraocular lenses. Current status and limitations].

Authors:  H B Dick; M Tehrani
Journal:  Ophthalmologe       Date:  2004-03       Impact factor: 1.059

3.  An interesting case of implantable contact lens.

Authors:  V K Mohindra; Savio Pereira
Journal:  Med J Armed Forces India       Date:  2013-06-21

4.  Causes of elevated intraocular pressure following implantation of phakic intraocular lenses for myopia.

Authors:  Salem Almalki; Abdullah Abubaker; Nasser A Alsabaani; Deepak P Edward
Journal:  Int Ophthalmol       Date:  2015-08-12       Impact factor: 2.031

Review 5.  [Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].

Authors:  T Kohnen; T Kasper; E Terzi
Journal:  Ophthalmologe       Date:  2005-11       Impact factor: 1.059

Review 6.  [Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].

Authors:  T Kohnen; M Baumeister; M Cichocki
Journal:  Ophthalmologe       Date:  2005-10       Impact factor: 1.059

Review 7.  [Scheimpflug photography for the examination of phakic intraocular lenses].

Authors:  M Baumeister
Journal:  Ophthalmologe       Date:  2014-10       Impact factor: 1.059

Review 8.  [Phakic intraocular lenses].

Authors:  T Kohnen; M Shajari
Journal:  Ophthalmologe       Date:  2016-06       Impact factor: 1.059

9.  Comparison of clear lens extraction and collamer lens implantation in high myopia.

Authors:  Ahmed M Emarah; Mostafa A El-Helw; Hazem M Yassin
Journal:  Clin Ophthalmol       Date:  2010-05-14

10.  Recurrent occlusion of laser iridotomy sites after posterior chamber phakic IOL implantation.

Authors:  In Ki Park; Je Myung Lee; Yeoun Sook Chun
Journal:  Korean J Ophthalmol       Date:  2008-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.