Literature DB >> 12683511

Pseudo-outbreak of Pseudomonas aeruginosa and Serratia marcescens related to bronchoscopes.

Cláudia V Silva1, Vanda D Magalhães, Crésio R Pereira, Julia Y Kawagoe, Chizue Ikura, Arnaldo J Ganc.   

Abstract

OBJECTIVE: To investigate an apparent outbreak involving simultaneous isolation of Pseudomonas aeruginosa and Serratia marcescens from bronchoalveolar lavage (BAL) samples.
DESIGN: Retrospective and prospective cohort studies using chart review, environmental sampling, and ribotyping of all available isolates. Cleaning and disinfection procedures for the bronchoscopes were also evaluated.
SETTING: A 380-bed private hospital in São Paulo, Brazil PATIENTS: Forty-one patients who underwent bronchoscopic procedures between December 1994 and October 1996 and from whom P. aeruginosa and S. marcescens were concomitantly isolated. Bronchoscopes and related items were microbiologically assessed.
RESULTS: P. aeruginosa and S. marcescens were simultaneously isolated from BAL samples 12.6% of the time (41 of 324) during the epidemic period versus 1.8% of the time (1 of 54) in the pre-epidemic period (P = .035). Ribotyping revealed two strains of P. aeruginosa and one of S. marcescens that were isolated from BAL samples of patients with no signs of respiratory tract infection, suggesting a pseudo-outbreak. Evaluation of bronchoscope disinfection revealed that inappropriate methods were being used. Implementation of simple control measures resulted in a significant decrease in simultaneous isolation of these species.
CONCLUSION: Prevention of pseudo-outbreaks requires meticulous use of preventive measures for infection-prone medical procedures.

Entities:  

Mesh:

Year:  2003        PMID: 12683511     DOI: 10.1086/502195

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  10 in total

1.  [Life-threatening necrotizing fasciitis colli caused by Serratia marcescens].

Authors:  M Meisel; H-J Schultz-Coulon
Journal:  HNO       Date:  2009-10       Impact factor: 1.284

2.  Pseudo-outbreak of extremely drug-resistant pseudomonas aeruginosa urinary tract infections due to contamination of an automated urine analyzer.

Authors:  M Hallin; A Deplano; S Roisin; V Boyart; R De Ryck; C Nonhoff; B Byl; Y Glupczynski; O Denis
Journal:  J Clin Microbiol       Date:  2012-01-04       Impact factor: 5.948

Review 3.  Serratia infections: from military experiments to current practice.

Authors:  Steven D Mahlen
Journal:  Clin Microbiol Rev       Date:  2011-10       Impact factor: 26.132

Review 4.  Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

Authors:  Julia Kovaleva; Frans T M Peters; Henny C van der Mei; John E Degener
Journal:  Clin Microbiol Rev       Date:  2013-04       Impact factor: 26.132

5.  Contaminated mouth swabs caused a multi-hospital outbreak of Pseudomonas aeruginosa infection.

Authors:  Bjørn G Iversen
Journal:  J Oral Microbiol       Date:  2010-04-20       Impact factor: 5.474

6.  Pseudomonas aeruginosa contamination of mouth swabs during production causing a major outbreak.

Authors:  Bjørn G Iversen; Hanne-Merete Eriksen; Gjermund Bø; Kristian Hagestad; Trond Jacobsen; Eva Engeset; Jørgen Lassen; Preben Aavitsland
Journal:  Ann Clin Microbiol Antimicrob       Date:  2007-03-13       Impact factor: 3.944

Review 7.  Infectious diseases linked to cross-contamination of flexible endoscopes.

Authors:  Nikki Kenters; Elisabeth G W Huijskens; Corianne Meier; Andreas Voss
Journal:  Endosc Int Open       Date:  2015-08

8.  Early Assessment of the Likely Cost Effectiveness of Single-Use Flexible Video Bronchoscopes.

Authors:  Christoffer Lilja Terjesen; Julia Kovaleva; Lars Ehlers
Journal:  Pharmacoecon Open       Date:  2017-06

9.  A systematic review and cost effectiveness analysis of reusable vs. single-use flexible bronchoscopes.

Authors:  J M Mouritsen; L Ehlers; J Kovaleva; I Ahmad; K El-Boghdadly
Journal:  Anaesthesia       Date:  2019-11-08       Impact factor: 6.955

10.  Evidence-based spectrum of antimicrobial activity for disinfection of bronchoscopes.

Authors:  Constanze Wendt; Birgit Kampf
Journal:  J Hosp Infect       Date:  2008-10       Impact factor: 3.926

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.