Literature DB >> 12683509

A decontamination and sterilization protocol employed during reuse of cardiac electrophysiology catheters inactivates human immunodeficiency virus.

Julian D Druce1, Jessica S Russell, Christopher J Birch, Lynette A Yates, Richard W Harper, Joseph J Smolich.   

Abstract

OBJECTIVE: To assess the effect of a standard decontamination and sterilization protocol employed during reuse of cardiac electrophysiology (EP) catheters on human immunodeficiency virus (HIV).
SETTING: Public health viral research laboratory.
METHODS: Studies were performed on distal, electrode-containing segments of 40 EP catheters previously used in up to 10 clinical EP procedures. EP catheter segments were immersed for 1 hour in blood contaminated with a high titer of HIV. After air drying for 2 hours, subgroups of 8 EP catheters were subjected to either (1) no treatment, (2) washing in general purpose detergent, (3) washing in enzyme cleaner, (4) sterilization in ethylene oxide, or (5) the full protocol of sequential detergent-enzyme cleaner-ethylene oxide exposure. HIV infectivity after treatment was determined by measuring HIV RNA and, in cell culture studies, assessing HIV-induced cytopathic effects (CPEs) and supernatant HIV-specific p24 antigen content
RESULTS: With no treatment, all catheters had high HIV RNA levels associated with CPEs and high p24 antigen levels. After washing in detergent, 5 of 8 catheters had HIV RNA detected, but without CPEs or p24 antigen. HIV RNA was detected in all catheters after washing in enzyme cleaner, with CPEs and a high p24 antigen level in 1 of 8 catheters. HIV RNA, CPEs, and p24 antigen were absent after ethylene oxide. After the full protocol, HIV RNA levels were undetectable (n = 7) or low (n = 1), without evidence of CPEs or p24 antigen.
CONCLUSION: Appropriate decontamination and sterilization of EP catheters during reuse is highly effective in inactivating HIV.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12683509     DOI: 10.1086/502190

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  4 in total

1.  Catheter interventions in congenital heart disease without regular catheterization laboratory equipment: the chain of hope experience in Rwanda.

Authors:  John Senga; Emmanuel Rusingiza; Joseph Mucumbitsi; Agnès Binagwaho; Bert Suys; Christine Lys; Karlien Carbonez; Caroline Ovaert; Thierry Sluysmans
Journal:  Pediatr Cardiol       Date:  2012-05-27       Impact factor: 1.655

2.  Catheter interventions for congenital heart disease in third world countries.

Authors:  R Krishna Kumar; M J Tynan
Journal:  Pediatr Cardiol       Date:  2005 May-Jun       Impact factor: 1.655

3.  Conception and validation of a protocol for reuse of non-irrigated electrophysiology catheters in a Brazilian teaching hospital.

Authors:  Mirtes Loeschner Leichsenring; Eliane Molina Psaltikidis; Márcio Jansen de Oliveira Figueiredo; Maria Luiza Moretti; Plínio Trabasso
Journal:  J Interv Card Electrophysiol       Date:  2017-12-11       Impact factor: 1.900

4.  Reprocessing of Medical Products in Electrophysiology.

Authors:  Ricardo Ryoshim Kuniyoshi; Eduardo Back Sternick; Elenir Nadalin; Denise Tessariol Hachul
Journal:  Arq Bras Cardiol       Date:  2017-02       Impact factor: 2.000

  4 in total