Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia.

N J Caplin1, P O'Leary, M Bulsara, E A Davis, T W Jones.   

Abstract

AIMS: To assess the accuracy of the Minimed continuous glucose monitoring system (CGMS) in estimating blood glucose concentration during a controlled reduction in blood glucose. RESEARCH DESIGN AND METHODS: We studied nine adolescent diabetics (age 14 +/- 1.5 years) wearing the CGMS during hyperinsulinaemic hypoglycaemic clamp studies. The glucose values obtained by the CGMS were compared with the venous blood samples taken during the studies and measured at the bedside using a glucose oxidase technique.
RESULTS: Blood glucose was lowered from euglycaemia to a mean of 2.8 mmol/l over 120 min and maintained at that level for a further 40 min. A total of 429 paired glucose measurements were available for analysis. Analysis using weighted Deming regression and t-tests revealed small differences between the methods, with blood glucose levels slightly higher than interstitial fluid levels. The mean difference across all values was 0.13 mmol/l. The observed difference was greatest at blood glucose values < 3.3 mmol/l.
CONCLUSIONS: This study suggests that during a fall in systemic glucose the subcutaneous glucose sensor provides an accurate reflection of blood glucose. However, a small difference is apparent at blood glucose values < 3.3 mmol/l.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12675670     DOI: 10.1046/j.1464-5491.2003.00837.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  21 in total

1.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 2.  A tale of two compartments: interstitial versus blood glucose monitoring.

Authors:  Eda Cengiz; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

3.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

4.  "Smart tattoo" glucose biosensors and effect of coencapsulated anti-inflammatory agents.

Authors:  Rohit Srivastava; Rahul Dev Jayant; Ayesha Chaudhary; Michael J McShane
Journal:  J Diabetes Sci Technol       Date:  2011-01-01

5.  Hypoglycemia impairs quality of blood glucose simulation in a clinical decision support system.

Authors:  Mette Dencker Johansen; Ole K Hejlesen; David A Cavan
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

6.  Continuous Glucose Monitoring After Gastric Bypass to Evaluate the Glucose Variability After a Low-Carbohydrate Diet and to Determine Hypoglycemia.

Authors:  Joan Bach Nielsen; Caroline Bruun Abild; Ane Mathilde Pedersen; Steen Bønløkke Pedersen; Bjørn Richelsen
Journal:  Obes Surg       Date:  2016-09       Impact factor: 4.129

7.  Glucose response of near-infrared alginate-based microsphere sensors under dynamic reversible conditions.

Authors:  Ayesha Chaudhary; Harri Harma; Pekka Hanninen; Michael J McShane; Rohit Srivastava
Journal:  Diabetes Technol Ther       Date:  2011-05-13       Impact factor: 6.118

8.  Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia.

Authors:  G M Steil; K Rebrin; F Hariri; S Jinagonda; S Tadros; C Darwin; M F Saad
Journal:  Diabetologia       Date:  2005-07-07       Impact factor: 10.122

9.  Esophageal and Gastric Dysmotilities are Associated with Altered Glucose Homeostasis and Plasma Levels of Incretins and Leptin.

Authors:  Rebecka Hammersjö; Bodil Roth; Peter Höglund; Bodil Ohlsson
Journal:  Rev Diabet Stud       Date:  2016-05-10

10.  Dissolved core alginate microspheres as "smart-tattoo" glucose sensors.

Authors:  Ayesha Chaudhary; Monica Raina; Michael J McShane; Rohit Srivastava
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.