Literature DB >> 12675591

Use of process data to assess chromatographic performance in production-scale protein purification columns.

Tina M Larson1, Jeff Davis, Harry Lam, Jerry Cacia.   

Abstract

Transition analysis was performed on production-scale chromatography data in order to monitor column performance. Analysis of over 300 transitions from several different chromatography operations demonstrated the utility of the techniques presented. Several of the transitions analyzed occurred on columns with known integrity breaches. The techniques proved sensitive for detection of these breaches. Seven transition calculations are presented, which were combined to produce a single overall integrity value for each column. In addition, principal components analysis (PCA) was used to detect shifts in the transition pattern, including those attributed to integrity breaches. Besides detection of integrity breaches, transition analysis proved useful in monitoring column efficiency over multiple column uses.

Mesh:

Substances:

Year:  2003        PMID: 12675591     DOI: 10.1021/bp025639g

Source DB:  PubMed          Journal:  Biotechnol Prog        ISSN: 1520-6033


  1 in total

1.  Packing of large-scale chromatography columns with irregularly shaped glass based resins using a stop-flow method.

Authors:  Sun Chau Siu; Celeste Chia; Yanglin Mok; Priyabrata Pattnaik
Journal:  Biotechnol Prog       Date:  2014-08-07
  1 in total