| Literature DB >> 12667677 |
Michiel C Warlé1, Jaap Kwekkeboom, Hugo W Tilanus, Herold J Metselaar.
Abstract
The use of the Immulite Automated Immunoassay system is becoming an increasingly popular method for accurate and fast measurement of sIL-2R in serum. Here, we report that detection of sIL-2R using the Immulite Immunoassay system in serum samples of patients who are treated with basiliximab, a monoclonal antibody against the alpha-chain of the IL-2 receptor, may lead to false measurements.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12667677 DOI: 10.1016/s0022-1759(03)00016-4