Heat effects on the retina.

Duke V Piyathaisere1, Eyal Margalit, Shih-Jen Chen, Jeng-Shyong Shyu, Salvatore A D'Anna, James D Weiland, Rhonda R Grebe, Lynnea Grebe, Gildo Fujii, Sahng Yeon Kim, Robert J Greenberg, Eugene De Juan, Mark S Humayun.   

Abstract

BACKGROUND AND
OBJECTIVE: To study the heat and power dissipation effect of anintraocular electronic heater on the retina. The determination of thermal parameters that are nonharmful to the retina will aid in the development of an implantable intraocular electronic retinal prosthesis.
MATERIALS AND METHODS: In dogs, five different retinal areas were touched with a custom intraocular heater probe (1.4 x 1.4 x 1.0 mm) for 1 second while the heater dissipated 0 (control), 10, 20, 50, or 100 mW. In a second protocol, the heater was mechanically held in the vitreous cavity while dissipating 500 mW for 2 hours while monitoring intraocular temperature. The animals were observed for 4 weeks with serial fundus photography and electroretinography. The procedure was then repeated in the fellow eye. The dogs were killed and both eyes were enucleated and submitted for histology.
RESULTS: In experiments using protocol 1, heater settings of 50 mW or higher caused an immediate visible whitening of the retinal tissue. Histologically, this damage was evident only if the eyeswere immediately enucleated. Permanent damage was caused by heater settings of 100 mW or higher. Under protocol 2, no ophthalmologic, electroretinography, or histologic differences were noted between the groups. Temperature increases of 5 degrees C in the vitreous and 2 degrees C near the retina were noted.
CONCLUSIONS: The liquid environment of the eye acts as a heat sink that is capable of dissipating a significant amount of power. An electronic chip positioned away from the retina can run at considerably higher powers than a chip positioned on the retinal surface.

Entities:  

Mesh:

Year:  2003        PMID: 12665226

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  7 in total

Review 1.  Wireless microstimulators for neural prosthetics.

Authors:  Mesut Sahin; Victor Pikov
Journal:  Crit Rev Biomed Eng       Date:  2011

2.  Visual Prosthesis: Artificial Vision.

Authors:  A Banarji; V S Gurunadh; S Patyal; T S Ahluwalia; D P Vats; M Bhadauria
Journal:  Med J Armed Forces India       Date:  2011-07-21

3.  THERMAL ANALYSIS OF INTRAOCULAR ELECTRONIC DISPLAY PROJECTOR VISUAL PROSTHESIS.

Authors:  D Gongal; S Thakur; A Panse; R Pawar; C Q Yu; C D Foster
Journal:  Numeri Heat Transf A Appl       Date:  2020-08-14       Impact factor: 2.928

4.  The development of neural stimulators: a review of preclinical safety and efficacy studies.

Authors:  Robert K Shepherd; Joel Villalobos; Owen Burns; David A X Nayagam
Journal:  J Neural Eng       Date:  2018-05-14       Impact factor: 5.379

5.  Investigation of thermal effects of infrared lasers on the rabbit retina: a study in the course of development of an active subretinal prosthesis.

Authors:  Heiko Sailer; Kei Shinoda; Georgios Blatsios; Konrad Kohler; Lars Bondzio; Eberhart Zrenner; Florian Gekeler
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-08       Impact factor: 3.535

Review 6.  Visual prostheses: the enabling technology to give sight to the blind.

Authors:  Mohammad Hossein Maghami; Amir Masoud Sodagar; Alireza Lashay; Hamid Riazi-Esfahani; Mohammad Riazi-Esfahani
Journal:  J Ophthalmic Vis Res       Date:  2014 Oct-Dec

Review 7.  Species Differences in the Nutrition of Retinal Ganglion Cells among Mammals Frequently Used as Animal Models.

Authors:  Christian Albrecht May
Journal:  Cells       Date:  2019-10-14       Impact factor: 6.600

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.