Literature DB >> 12665035

Noncontact radio-frequency ablation for obtaining deeper lesions.

Jie Zhang1, Jang-Zern Tsai, Hong Cao, Yi Chen, James A Will, Vicken R Vorperian, John G Webster.   

Abstract

Radio-frequency (RF) cardiac catheter ablation has been very successful for treating some cardiac arrhythmias, however, the success rate for ventricular tachycardias is still not satisfactory. Some existing methods for developing deeper lesions include active cooling of the electrode and modifying the electrode shape. We propose a method of noncontact ablation, to solve this problem. We apply 120 W of power through an 8-mm electrode for a 120-s duration, with distances from 0 to 3 mm between electrode and myocardium, to create lesions in myocardium. We apply flow rates of 1, 3, and 5 L/min to determine their effect. Results show that with an optimal distance from 0.5 to 1.5 mm between electrode and myocardium, we increase lesion depth from 7.5 mm for contact ablation to 9.5 mm for noncontact ablation. For different flow rates, the optimal distance various. The effect of flow rate is not obvious. Higher flow rate does not lead to a deeper lesion.

Entities:  

Mesh:

Year:  2003        PMID: 12665035     DOI: 10.1109/TBME.2002.807647

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

1.  Finite element analysis of hepatic radiofrequency ablation probes using temperature-dependent electrical conductivity.

Authors:  Isaac Chang
Journal:  Biomed Eng Online       Date:  2003-05-08       Impact factor: 2.819