Measurement of total homocysteine in plasma and blood spots using liquid chromatography-tandem mass spectrometry: comparison with the plasma Abbott IMx method.

Steven J McCann1, Scott Gillingwater, Brian G Keevil, Donald P Cooper, Michel R Morris.   

Abstract

BACKGROUND: Current sampling for total homocysteine (tHcy) is problematic, requiring plasma separation within 15 min. The aim of this study was to develop a liquid chromatographic-tandem mass spectrometric (LC-MS/MS) method for the measurement of tHcy in plasma and dried blood spots and to determine whether the dried blood spot concentration could be used to predict plasma concentrations of tHcy.
METHODS: LC-MS/MS methodology was optimized to measure tHcy in plasma and dried blood spots. Fifty blood samples collected from heart transplant patients were used to form dried blood spots and for plasma analysis. Plasma tHcy was also measured using the Abbott IMx method and values were compared to the tHcy concentrations determined in plasma and dried blood spots using LC-MS/MS methodology.
RESULTS: The plasma tHcy LC-MS/MS results compared well with the IMx values: LC-MS/MS=1.18(IMx)-0.44 (r(2)=0.915). The within-batch precision (n =10) of the plasma LC-MS/MS method was < 2.0% at 14.6 and 37.7 micromol/L, respectively; the between-batch precision (n=10) was 5.0 and 8.0%, respectively, at these concentrations. The method was found to be sensitive down to 1 micromol/L and linear up to at least 100 micromol/L. Dried blood spot LC-MS/MS results were considerably lower than the plasma IMx values (P < 0.0001): dried blood spot LC-MS/MS=0.33IMx+1.77 (r(2)=0.682). The within-batch precision (n=20) of the dried blood spot LC-MS/MS method was 7.3% and 4.7% at concentrations of 4.0 and 7.9 micromol/L, respectively; the between-batch precision was 12.6% and 7.9% at concentrations of 5.1 and 8.0 micromol/L, respectively. To assess whether dried blood spots are suitable as a screening test to predict plasma tHcy concentrations, arbitary cut-off levels were compared. If it is assumed that a plasma tHcy concentration of >15 micromol/L is raised, a dried blood spot result of >6.8 micro mol/L has a sensitivity and specificity in detecting a raised plasma tHcy of 83.3% and 96.2%, respectively, and a positive and negative predictive value of 95% and 86%, respectively, with an efficiency of 90%. Use of a dried blood spot cut-off concentration of 6.2 micromol/L for predicting high plasma tHcy concentrations (above 15 micromol/L) has a sensitivity and specificity of 95.8% and 73.1%, respectively, positive and negative predictive values of 76% and 95%, respectively, and an efficiency of 84%.
CONCLUSIONS: We have developed a precise and accurate LC-MS/MS method for measuring plasma tHcy concentrations, which uses a small volume of plasma and is suitable for routine use. A satisfactory LC-MS/MS method for the measurement of tHcy in dried blood spots was also developed; this method might be useful in routine screening for raised plasma concentrations of tHcy.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12662404     DOI: 10.1258/000456303763046094

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  9 in total

Review 1.  What a drop can do: dried blood spots as a minimally invasive method for integrating biomarkers into population-based research.

Authors:  Thomas W McDade; Sharon Williams; J Josh Snodgrass
Journal:  Demography       Date:  2007-11

2.  Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry.

Authors:  James J Pitt
Journal:  Clin Biochem Rev       Date:  2009-02

3.  Homocysteine measurement in dried blood spot for neonatal detection of homocystinurias.

Authors:  Ahmad N Alodaib; Kevin Carpenter; Veronica Wiley; Tiffany Wotton; John Christodoulou; Bridget Wilcken
Journal:  JIMD Rep       Date:  2011-12-13

4.  Gene-environment interactions reveal a homeostatic role for cholesterol metabolism during dietary folate perturbation in mice.

Authors:  Toshimori Kitami; Renee Rubio; William O'Brien; John Quackenbush; Joseph H Nadeau
Journal:  Physiol Genomics       Date:  2008-08-12       Impact factor: 3.107

Review 5.  Guidelines for the diagnosis and management of cystathionine beta-synthase deficiency.

Authors:  Andrew A M Morris; Viktor Kožich; Saikat Santra; Generoso Andria; Tawfeg I M Ben-Omran; Anupam B Chakrapani; Ellen Crushell; Mick J Henderson; Michel Hochuli; Martina Huemer; Miriam C H Janssen; Francois Maillot; Philip D Mayne; Jenny McNulty; Tara M Morrison; Helene Ogier; Siobhan O'Sullivan; Markéta Pavlíková; Isabel Tavares de Almeida; Allyson Terry; Sufin Yap; Henk J Blom; Kimberly A Chapman
Journal:  J Inherit Metab Dis       Date:  2016-10-24       Impact factor: 4.982

6.  A disrupted transsulphuration pathway results in accumulation of redox metabolites and induction of gametocytogenesis in malaria.

Authors:  Divya Beri; Balu Balan; Shweta Chaubey; Suraj Subramaniam; Bachu Surendra; Utpal Tatu
Journal:  Sci Rep       Date:  2017-01-16       Impact factor: 4.379

7.  Advantages and Challenges of Dried Blood Spot Analysis by Mass Spectrometry Across the Total Testing Process.

Authors:  Rosita Zakaria; Katrina J Allen; Jennifer J Koplin; Peter Roche; Ronda F Greaves
Journal:  EJIFCC       Date:  2016-12-01

8.  New Developments in Biomarkers for Atopic Dermatitis.

Authors:  Judith L Thijs; Wouter van Seggelen; Carla Bruijnzeel-Koomen; Marjolein de Bruin-Weller; DirkJan Hijnen
Journal:  J Clin Med       Date:  2015-03-16       Impact factor: 4.241

9.  Analysis of serum homocysteine in the laboratory practice - comparison of the direct chemiluminescence immunoassay and high performance liquid chromatography coupled with fluorescent detection.

Authors:  Łukasz Paprotny; Dorota Wianowska; Magdalena Izdebska; Agnieszka Celejewska; Dorota Szewczak; Janusz Solski
Journal:  Biochem Med (Zagreb)       Date:  2020-08-05       Impact factor: 2.313

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.