Literature DB >> 12661825

Performance evaluation of the Sysmex XT-2000i automated hematology analyzer.

K Langford1, L Luchtman-Jones, R Miller, D Walck.   

Abstract

The Sysmex XT-2000i automated hematology analyzer was evaluated at Saint Louis Children's Hospital (SLCH), St. Louis, MO, USA. Complete blood count results from the Sysmex XT-2000i were compared to results from the Sysmex XE-2100 for 114 pediatric and adult patient samples. Manual differentials were performed on each specimen by 2 experienced medical technologists using guidelines established in the National Committee for Clinical Laboratory Standards (NCCLS) document H20-A. Carryover, precision, linearity, correlation, stability, and mixing-test studies were also performed. The XT-2000i results showed excellent correlation with the results from the XE-2100 for the following parameters: white blood cells; red blood cells; hemoglobin; hematocrit; mean corpuscular volume; mean corpuscular hemoglobin; mean corpuscular hemoglobin concentration; red blood cell distribution width by standard deviation; red blood cell distribution width by coefficient of variation; mean platelet volume; platelets; percent neutrophils, lymphocytes, monocytes, eosinophils, and basophils; and reticulocyte percent and number. In our evaluation of the XT-2000i the correlation coefficients for all complete blood counts and differential parameters compared well with those of the XE-2100. We concluded that the XT-2000i demonstrated comparable analytical performance to its predecessor, the XE-2100.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12661825

Source DB:  PubMed          Journal:  Lab Hematol        ISSN: 1080-2924


  2 in total

1.  Prognostic value of hemoglobin decline over the GRACE score in patients hospitalized for an acute coronary syndrome.

Authors:  Sylvestre Maréchaux; Stéphanie Barrailler; Claire Pinçon; Vincent Decourcelle; Thomas Guidez; Sophie Braun; Nadia Bouabdallaoui; Jean-Jacques Bauchart; Jean Luc Auffray; Francis Juthier; Carlo Banfi; Sophie Susen; Brigitte Jude; Philippe Asseman; Eric Van Belle; Pierre Vladimir Ennezat
Journal:  Heart Vessels       Date:  2011-05-12       Impact factor: 2.037

2.  Evaluation of the Performance of the Sysmex XT-2000i Hematology Analyzer With Whole Bloods Stored at Room Temperature.

Authors:  Valerie L Hill; Virginia Z Simpson; Jeanette M Higgins; Zonghui Hu; Randy A Stevens; Julie A Metcalf; Michael Baseler
Journal:  Lab Med       Date:  2009-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.