Impact of target volume coverage with Radiation Therapy Oncology Group (RTOG) 98-05 guidelines for transrectal ultrasound guided permanent Iodine-125 prostate implants.

Eric M Horwitz1, Raj K Mitra, Robert G Uzzo, Indra J Das, Wayne H Pinover, Alexandra L Hanlon, Shawn W McNeeley, Gerald E Hanks.   

Abstract

PURPOSE: Despite the wide use of permanent prostate implants for the treatment of early stage prostate cancer, there is no consensus for optimal pre-implant planning guidelines that results in maximal post-implant target coverage. The purpose of this study was to compare post-implant target volume coverage and dosimetry between patients treated before and after Radiation Therapy Oncology Group (RTOG) 98-05 guidelines were adopted using several dosimetric endpoints.
MATERIALS AND METHODS: Ten consecutively treated patients before the adoption of the RTOG 98-05 planning guidelines were compared with ten consecutively treated patients after implementation of the guidelines. Pre-implant planning for patients treated pre-RTOG was based on the clinical target volume (CTV) defined by the pre-implant TRUS definition of the prostate. The CTV was expanded in each dimension according to RTOG 98-05 and defined as the planning target volume. The evaluation target volume was defined as the post-implant computed tomography definition of the prostate based on RTOG 98-05 protocol recommendations. Implant quality indicators included V(100), V(90), V(100), and Coverage Index (CI).
RESULTS: The pre-RTOG median V(100), V(90), D(90), and CI values were 82.8, 88.9%, 126.5 Gy, and 17.1, respectively. The median post-RTOG V(100), V(90), D(90), and CI values were 96.0, 97.8%, 169.2 Gy, and 4.0, respectively. These differences were all statistically significant.
CONCLUSIONS: Implementation of the RTOG 98-05 implant planning guidelines has increased coverage of the prostate by the prescription isodose lines compared with our previous technique, as indicated by post-implant dosimetry indices such as V(100), V(90), D(90). The CI was also improved significantly with the protocol guidelines. Our data confirms the validity of the RTOG 98-05 implant guidelines for pre-implant planning as it relates to enlargement of the CTV to ensure adequate margin between the CTV and the prescription isodose lines.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12648789     DOI: 10.1016/s0167-8140(02)00387-0

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

1.  A magnetic resonance-based seed localization method for I-125 prostate implants.

Authors:  Rena J Lee; Hyun Suk Suh; Kyung Ja Lee; Soome Lim; Yookyung Kim; Sungkyu Kim; Jinho Choi
Journal:  J Korean Med Sci       Date:  2007-09       Impact factor: 2.153

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.