Literature DB >> 12641610

Automated detection of malaria-associated intraleucocytic haemozoin by Cell-Dyn CD4000 depolarization analysis.

C S Scott1, D van Zyl, E Ho, D Meyersfeld, L Ruivo, B V Mendelow, T L Coetzer.   

Abstract

Laboratory tests for malaria are only performed if there is clinical suspicion of the disease, and a missed diagnosis contributes substantially to morbidity and mortality. Malaria parasites produce haemozoin, which is able to depolarize light and this allows the automated detection of malaria during routine complete blood count analysis (CBC) with some Abbott Cell-Dyn instruments. In this study, we evaluated the Cell-Dyn CD4000 with 831 blood samples submitted for malaria investigations. Samples were categorized as malaria negative (n = 417), convalescent malaria (n = 64) or malaria positive (n = 350) by reference to thin/thick film microscopy, 'rapid test' procedures, polymerase chain reaction analysis and clinical history. With regard to CD4000 depolarization analysis, a malaria positive CD4000 pattern was ascribed to samples that showed one or more abnormal depolarizing purple events, which corresponded to monocytes containing ingested malaria pigment (haemozoin). Positive CD4000 patterns were observed in 11 of 417, 50 of 64 and 281 of 350 of malaria negative, convalescent malaria and malaria positive samples respectively. The specificity and positive predictive values for malaria (active and convalescent) were very high (97.4 and 96.8%, respectively), while sensitivity and negative predictive values were 80.0 and 83.0% respectively. Depolarization analysis was particularly effective for Plasmodium falciparum malaria but there was lower detection sensitivity for White compared with Black African patients. CD4000 90 degrees depolarization vs 0 degrees analysis revealed a proportion of samples with small nonleucocyte-associated depolarizing particles. Appearance of such events in the form of a discrete cluster was associated with P. vivax rather than P. falciparum infection.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12641610     DOI: 10.1046/j.1365-2257.2003.00496.x

Source DB:  PubMed          Journal:  Clin Lab Haematol        ISSN: 0141-9854


  7 in total

1.  Application of Sysmex XN-Series Automated Haematology Analyser in the Rapid Detection of Malaria.

Authors:  Aparna Ningombam; Abhirup Sarkar; Shreyam Acharya; Anita Chopra; Kundan Kumar; Arulselvi Subramanian
Journal:  Indian J Hematol Blood Transfus       Date:  2020-04-03       Impact factor: 0.900

2.  Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area.

Authors:  Jong-Ha Yoo; Jaewoo Song; Kyung-A Lee; Young-Kyu Sun; Young-Ah Kim; Tae Sung Park; Jong Rak Choi
Journal:  Am J Trop Med Hyg       Date:  2010-03       Impact factor: 2.345

3.  Is flow cytometry better in counting malaria pigment-containing leukocytes compared to microscopy?

Authors:  Thomas Hänscheid; Rosangela Frita; Matthias Längin; Peter G Kremsner; Martin P Grobusch
Journal:  Malar J       Date:  2009-11-16       Impact factor: 2.979

Review 4.  A research agenda for malaria eradication: diagnoses and diagnostics.

Authors: 
Journal:  PLoS Med       Date:  2011-01-25       Impact factor: 11.069

Review 5.  Automated haematology analysis to diagnose malaria.

Authors:  Germán Campuzano-Zuluaga; Thomas Hänscheid; Martin P Grobusch
Journal:  Malar J       Date:  2010-11-30       Impact factor: 2.979

6.  Full blood count and haemozoin-containing leukocytes in children with malaria: diagnostic value and association with disease severity.

Authors:  Thomas Hänscheid; Matthias Längin; Bertrand Lell; Marc Pötschke; Sunny Oyakhirome; Peter G Kremsner; Martin P Grobusch
Journal:  Malar J       Date:  2008-06-12       Impact factor: 2.979

7.  A diagnostic tool for malaria based on computer software.

Authors:  Manas Kotepui; Kwuntida Uthaisar; Bhukdee Phunphuech; Nuoil Phiwklam
Journal:  Sci Rep       Date:  2015-11-12       Impact factor: 4.379

  7 in total