Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients.

Andreas Bur1, Harald Herkner, Marianne Vlcek, Christian Woisetschläger, Ulla Derhaschnig, Georg Delle Karth, Anton N Laggner, Michael M Hirschl.   

Abstract

OBJECTIVE: Comparison of oscillometric blood pressure measurement with two different devices (M3000A using a new algorithm and M1008A using an established algorithm, both Hewlett Packard) and evaluation of current recommendations concerning the relation between cuff size and upper arm circumference in critically ill patients.
DESIGN: Prospective data collection.
SETTING: Emergency department in a 2000-bed inner-city hospital. PATIENTS: A total of 30 patients categorized into three groups according to their upper arm circumference (I, 18-25 cm; II, 25.1-33 cm; III, 33.1-47.5 cm) were enrolled in the study protocol. INTERVENTIONS In each patient, two noninvasive blood pressure devices with three different cuff sizes were used to perform oscillometric blood pressure measurement. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery. MEASUREMENT AND MAIN
RESULTS: Overall, 1,011 pairs of simultaneous oscillometric and invasive blood pressure measurements were collected in 30 patients (group I, n = 10; group II, n = 10; group III, n = 10). The overall discrepancy between both methods with the M3000A was -2.4 +/- 11.8 mm Hg (p <.0001) and, with the M1008A, -5.3 +/- 11.6 mm Hg (p <.0001) if the recommended cuff size according to the upper arm circumference was used (352 measurements). If smaller cuff sizes than recommended were used (308 measurements performed in group II and III), the overall discrepancy between both methods with the M3000A was 1.3 +/- 13.4 mm Hg (p <.024) and, with the M1008A, -2.3 +/- 11.5 mm Hg (p <.0001).
CONCLUSION: The new algorithm reduced the overall bias of the oscillometric method but still showed a significant discrepancy between both methods of blood pressure measurement, primarily due to the mismatch between upper arm circumference and cuff size. The improvement of the algorithm alone could not result in a sufficient improvement of oscillometric blood pressure measurement. A reevaluation of the recommendations concerning the relation between upper arm circumference and cuff size are urgently required if oscillometric blood pressure measurement should become a reasonable alternative to intra-arterial blood pressure measurement in critically ill patients.

Entities:  

Mesh:

Year:  2003        PMID: 12626986     DOI: 10.1097/01.CCM.0000053650.12025.1A

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  24 in total

1.  The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2010-06-08       Impact factor: 2.502

2.  Resonance artefacts in modern pressure monitoring systems.

Authors:  L Bocchi; S Romagnoli
Journal:  J Clin Monit Comput       Date:  2015-08-27       Impact factor: 2.502

3.  Changes in pulse pressure following fluid loading: a comparison between aortic root (non-invasive tonometry) and femoral artery (invasive recordings).

Authors:  Nicolas Dufour; Denis Chemla; Jean-Louis Teboul; Xavier Monnet; Christian Richard; David Osman
Journal:  Intensive Care Med       Date:  2011-03-05       Impact factor: 17.440

Review 4.  Non-invasive continuous blood pressure monitoring: a review of current applications.

Authors:  Elena Chung; Guo Chen; Brenton Alexander; Maxime Cannesson
Journal:  Front Med       Date:  2013-01-23       Impact factor: 4.592

5.  Photoplethysmographic characterization of vascular tone mediated changes in arterial pressure: an observational study.

Authors:  Gerardo Tusman; Cecilia M Acosta; Sven Pulletz; Stephan H Böhm; Adriana Scandurra; Jorge Martinez Arca; Matías Madorno; Fernando Suarez Sipmann
Journal:  J Clin Monit Comput       Date:  2018-12-15       Impact factor: 2.502

Review 6.  Home Blood Pressure Monitoring.

Authors:  Jacob George; Thomas MacDonald
Journal:  Eur Cardiol       Date:  2015-12

7.  Correlation of Noninvasive Blood Pressure and Invasive Intra-arterial Blood Pressure in Patients Treated with Vasoactive Medications in a Neurocritical Care Unit.

Authors:  Ali A Saherwala; Sonja E Stutzman; Mohamed Osman; Junaid Kalia; Stephen A Figueroa; DaiWai M Olson; Venkatesh Aiyagari
Journal:  Neurocrit Care       Date:  2018-06       Impact factor: 3.210

Review 8.  Therapeutic hypertension: principles and methods.

Authors:  David J Powner; Joseph M Darby; John W Crommett; Robert L Levine
Journal:  Neurosurg Rev       Date:  2004-08-14       Impact factor: 3.042

9.  Variation of arterial and central venous catheter use in United States intensive care units.

Authors:  Hayley B Gershengorn; Allan Garland; Andrew Kramer; Damon C Scales; Gordon Rubenfeld; Hannah Wunsch
Journal:  Anesthesiology       Date:  2014-03       Impact factor: 7.892

10.  The mean machine; accurate non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2009-08-12       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.