Reproducibility and clinical application of a newly developed stabilized retinal laser Doppler instrument.

Akitoshi Yoshida1, Gilbert T Feke, Fumihiko Mori, Taiji Nagaoka, Naoki Fujio, Hironobu Ogasawara, Suguru Konno, J Wallace Mcmeel.   

Abstract

PURPOSE: To describe a newly developed stabilized retinal laser Doppler instrument, to report the reproducibility of retinal blood flow measurements, and to present examples of its clinical application.
DESIGN: Experimental study.
METHODS: The intrasession, intersession, and interobserver reproducibility of retinal blood flow measurements obtained using the Canon Laser Blood Flowmeter model CLBF 100 was assessed. Intrasession: the coefficients of variation (CV) for repeated measurements (5X) of retinal vessel diameter (D), time-average centerline blood velocity (V(av)), and blood flow (F) were calculated at 18 sites along temporal retinal arteries and 18 sites along temporal retinal veins using both eyes in six healthy volunteers. Intersession: the correlation coefficients and average differences between two sets of measurements at the same 36 retinal sites made by one examiner on two different days were calculated. Interobserver: the correlation coefficients and average differences between two sets of measurements at nine sites (five arterial and four venous) in five eyes of five of the volunteers made by two examiners on the same day were calculated. Results from two patients, one with branch retinal vein occlusion (BRVO), and one with central retinal vein occlusion (CRVO), are used to illustrate abnormal retinal circulatory characteristics.
RESULTS: intrasession: CV (mean +/- SD) for D, V(av), and F were 5.5% +/- 2.2%, 13.1% +/- 6.3%, and 13.8% +/- 4.7% in arteries and 3.5% +/- 2.1%, 11.9% +/- 5.9%, and 12.7% +/- 5.5% in veins. Intersession/interobserver: strong correlations between measurements made by one examiner on two different days and by two examiners on the same day were found in D, V(av), and F. Average differences in F were 16.4% +/- 12.8% for intersession and 12.6% +/- 7.0% for interobserver comparisons. Results from the patient with BRVO show marked differences in flow characteristics in the affected quadrant compared with an unaffected quadrant in the same eye, and to the fellow eye. Results from the patient with CRVO show a dramatic improvement in flow characteristics after clearing of the occlusion.
CONCLUSIONS: The reproducibility results and the fact that blood flow is measured in actual units of microl/min indicate that the instrument can be used for reliable comparison of blood flow characteristics at different retinal vascular sites in the same eye, at comparable sites in both eyes, and for comparison between patients and healthy control subjects. Copyright 2003 by Elsevier Science Inc.

Entities:  

Mesh:

Year:  2003        PMID: 12614754     DOI: 10.1016/s0002-9394(02)01949-9

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  22 in total

Review 1.  Retinal vascular image analysis as a potential screening tool for cerebrovascular disease: a rationale based on homology between cerebral and retinal microvasculatures.

Authors:  Niall Patton; Tariq Aslam; Thomas Macgillivray; Alison Pattie; Ian J Deary; Baljean Dhillon
Journal:  J Anat       Date:  2005-04       Impact factor: 2.610

2.  Validation of spectral domain optical coherence tomographic Doppler shifts using an in vitro flow model.

Authors:  Larry Kagemann; Gadi Wollstein; Hiroshi Ishikawa; Kelly A Townsend; Joel S Schuman
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-09-29       Impact factor: 4.799

3.  Reliability of vessel diameter measurements with a retinal oximeter.

Authors:  Renata Blondal; Margret Kara Sturludottir; Sveinn Hakon Hardarson; Gisli Hreinn Halldorsson; Einar Stefánsson
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-04-16       Impact factor: 3.117

4.  Conjunctival microvascular haemodynamics in sickle cell retinopathy.

Authors:  Ali Kord Valeshabad; Justin Wanek; Ruth Zelkha; Jennifer I Lim; Nicole Camardo; Bruce Gaynes; Mahnaz Shahidi
Journal:  Acta Ophthalmol       Date:  2014-11-27       Impact factor: 3.761

5.  Effects of topical carteolol on retinal arterial blood flow in primary open-angle glaucoma patients.

Authors:  Motofumi Kawai; Taiji Nagaoka; Atsushi Takahashi; Eiichi Sato; Akitoshi Yoshida
Journal:  Jpn J Ophthalmol       Date:  2012-06-14       Impact factor: 2.447

6.  Effect of haemodialysis on retinal circulation in patients with end stage renal disease.

Authors:  T Nagaoka; Y Takeyama; S Kanagawa; K Sakagami; F Mori; A Yoshida
Journal:  Br J Ophthalmol       Date:  2004-08       Impact factor: 4.638

7.  Changes in central retinal artery blood flow after ocular warming and cooling in healthy subjects.

Authors:  M A Shamshad; A K Amitava; I Ahmad; S Wahab
Journal:  Indian J Ophthalmol       Date:  2010 May-Jun       Impact factor: 1.848

8.  Segmental reproducibility of retinal blood flow velocity measurements using retinal function imager.

Authors:  Jay Chhablani; Dirk-Uwe Bartsch; Lingyun Cheng; Laura Gomez; Rayan A Alshareef; Sami S Rezeq; Sunir J Garg; Zvia Burgansky-Eliash; William R Freeman
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-05-23       Impact factor: 3.117

9.  Reduction of retinal blood flow in high myopia.

Authors:  Noriaki Shimada; Kyoko Ohno-Matsui; Seiyo Harino; Takeshi Yoshida; Kenjiro Yasuzumi; Ariko Kojima; Kanako Kobayashi; Soh Futagami; Takashi Tokoro; Manabu Mochizuki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-01-14       Impact factor: 3.117

10.  Pilot study of Doppler optical coherence tomography of retinal blood flow following laser photocoagulation in poorly controlled diabetic patients.

Authors:  Jennifer C Lee; Brandon J Wong; Ou Tan; Sowmya Srinivas; Srinivas R Sadda; David Huang; Amani A Fawzi
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-09-09       Impact factor: 4.799

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.