Comparison of ankle-brachial pressure index measurements using an automated oscillometric device with the standard Doppler ultrasound technique.

Anantha Ramanathan1, Philip J Conaghan, Andrew D Jenkinson, Christopher R Bishop.   

Abstract

BACKGROUND: The aim of the present study was to assess the accuracy of an automated oscillometric device (DINAMAP) in measuring ankle-brachial pressure index (ABPI) in the clinical setting. Results were then compared to those obtained using the standard Doppler technique.
METHODS: A prospective comparative study of 50 healthy volunteers was carried out using a standard ward DINAMAP machine, hand-held Doppler and a 14 cm sphygmomanometer cuff. Three consecutive pressure readings were taken from all four limbs of the volunteers using both Doppler and DINAMAP techniques. The two techniques were carried out by different investigators who were blind to the findings of the other. The ABPI was calculated for each technique and the two techniques were compared.
RESULTS: There was no significant agreement between the ABPI readings obtained by DINAMAP and those obtained by the standard Doppler technique. There was also a substantial variation in pressure readings produced by the DINAMAP for a given limb as measured by intraclass correlation of the three readings taken.
CONCLUSIONS: There is no role for standard automated oscillometric devices in the calculation of ABPI in the vascular clinic.

Mesh:

Year:  2003        PMID: 12608969     DOI: 10.1046/j.1445-2197.2003.02582.x

Source DB:  PubMed          Journal:  ANZ J Surg        ISSN: 1445-1433            Impact factor:   1.872


  5 in total

1.  Are automated blood pressure monitors accurate enough to calculate the ankle brachial pressure index?

Authors:  Euan MacDonald; Paul Froggatt; Gwen Lawrence; Stephen Blair
Journal:  J Clin Monit Comput       Date:  2008-11-06       Impact factor: 2.502

2.  Ankle-Brachial index by oscillometry: A very useful method to assess peripheral arterial disease in diabetes.

Authors:  M Premanath; M Raghunath
Journal:  Int J Diabetes Dev Ctries       Date:  2010-04

3.  Interrater and intrarater reliability of photoplethysmography for measuring toe blood pressure and toe-brachial index in people with diabetes mellitus.

Authors:  Christopher Scanlon; Kris Park; David Mapletoft; Lindy Begg; Joshua Burns
Journal:  J Foot Ankle Res       Date:  2012-06-07       Impact factor: 2.303

4.  The reliability of the ankle-brachial index in the Atherosclerosis Risk in Communities (ARIC) study and the NHLBI Family Heart Study (FHS).

Authors:  Beth D Weatherley; Lloyd E Chambless; Gerardo Heiss; Diane J Catellier; Curtis R Ellison
Journal:  BMC Cardiovasc Disord       Date:  2006-02-21       Impact factor: 2.298

5.  The reliability of the ankle brachial index: a systematic review.

Authors:  Sarah Casey; Sean Lanting; Christopher Oldmeadow; Vivienne Chuter
Journal:  J Foot Ankle Res       Date:  2019-08-02       Impact factor: 2.303

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.