A new rapid threshold algorithm for short-wavelength automated perimetry.

Boel Bengtsson1.   

Abstract

PURPOSE: To develop and test a short and reliable visual field threshold program for the early detection of glaucomatous visual field loss, by adapting the Swedish interactive test algorithm (SITA) to short-wavelength-automated-perimetry (SWAP).
METHODS: Computer simulations were performed to test the accuracy of several versions of SITA SWAP, and to optimize speed versus reliability. The selected SITA SWAP version was evaluated and compared with the older Full Threshold SWAP and Fastpac SWAP programs in 41 patients with glaucoma and normal subjects.
RESULTS: Average test time was 3.6 minutes for SITA SWAP, 11.8 minutes for Full Threshold SWAP, and 7.7 minutes for Fastpac SWAP, differences were significant at P < 0.0001. Mean threshold reproducibility, calculated as absolute difference between two tests, did not differ significantly between programs and was 2.4 dB for SITA, 2.3 dB for Full Threshold, and 2.4 dB for Fastpac SWAP. Simultaneous comparison showed significant differences in threshold sensitivity, P = 0.023: SITA SWAP showed highest sensitivity, 21.6 dB on average, compared with both Full Threshold SWAP and Fastpac SWAP with a mean sensitivity of 17.3 and 17.8 dB, respectively.
CONCLUSIONS: SITA SWAP was much faster than the older SWAP strategies, and reproducibility did not differ. This implies that SITA SWAP could become a clinically useful method for the detection of early glaucoma. SWAP tests may also be applicable in larger groups of patients because of the increased dynamic range.

Entities:  

Mesh:

Year:  2003        PMID: 12601072     DOI: 10.1167/iovs.02-0169

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  9 in total

1.  African Descent and Glaucoma Evaluation Study (ADAGES): III. Ancestry differences in visual function in healthy eyes.

Authors:  Lyne Racette; Jeffrey M Liebmann; Christopher A Girkin; Linda M Zangwill; Sonia Jain; Lida M Becerra; Felipe A Medeiros; Christopher Bowd; Robert N Weinreb; Catherine Boden; Pamela A Sample
Journal:  Arch Ophthalmol       Date:  2010-05

2.  Visual fields correlate better than visual acuity to severity of diabetic retinopathy.

Authors:  B Bengtsson; A Heijl; E Agardh
Journal:  Diabetologia       Date:  2005-11-01       Impact factor: 10.122

3.  Comparing the full-threshold and Swedish interactive thresholding algorithms for short-wavelength automated perimetry.

Authors:  Minna Ng; Lyne Racette; John P Pascual; Jeffrey M Liebmann; Christopher A Girkin; Sarah L Lovell; Linda M Zangwill; Robert N Weinreb; Pamela A Sample
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-12-13       Impact factor: 4.799

4.  Assessment of patient opinions of different clinical tests used in the management of glaucoma.

Authors:  Stuart K Gardiner; Shaban Demirel
Journal:  Ophthalmology       Date:  2008-12       Impact factor: 12.079

Review 5.  Diagnostic tools for glaucoma detection and management.

Authors:  Pooja Sharma; Pamela A Sample; Linda M Zangwill; Joel S Schuman
Journal:  Surv Ophthalmol       Date:  2008-11       Impact factor: 6.048

6.  Visual function-specific perimetry to identify glaucomatous visual loss using three different definitions of visual field abnormality.

Authors:  Ali Tafreshi; Pamela A Sample; Jeffrey M Liebmann; Christopher A Girkin; Linda M Zangwill; Robert N Weinreb; Maziar Lalezary; Lyne Racette
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-10-31       Impact factor: 4.799

7.  Long-term variability in short-wavelength automated perimetry compared to standard perimetry in glaucoma patients and normal subjects.

Authors:  Chintan Malhotra
Journal:  Oman J Ophthalmol       Date:  2009-01

8.  A Method for Reducing the Number of Presentations in Perimetric Test Procedures.

Authors:  Andrew Turpin; Allison M McKendrick
Journal:  Transl Vis Sci Technol       Date:  2022-04-01       Impact factor: 3.048

9.  Comparison of Standard Automated Perimetry, Short-Wavelength Automated Perimetry, and Frequency-Doubling Technology Perimetry to Monitor Glaucoma Progression.

Authors:  Rongrong Hu; Chenkun Wang; Yangshun Gu; Lyne Racette
Journal:  Medicine (Baltimore)       Date:  2016-02       Impact factor: 1.889