| Literature DB >> 1260032 |
H L Schmidt, G Krisam, G Grenner.
Abstract
The enthalpy of processes catalyzed by immobilized enzymes in the reaction cell of a LKB-flow calorimeter is used for determination of urea (0.5-5 mumol) and glucose (0.03-0.5 mumol). Accuracy is 2-5% and the time needed for one analysis is 20 min. A sensitive "enzyme thermistor" consisting of a flow through cell with an immobilized enzyme and two thermistors is described, which permits glucose determinations (0.05-1 mumol +/- 0.03 mumol) by means of temperature difference caused by reaction heat. Coupling of enzyme reactions for increasing reaction heat and consequently sensitivity in calorimetric determinations is demonstrated.Entities:
Mesh:
Substances:
Year: 1976 PMID: 1260032 DOI: 10.1016/0005-2744(76)90051-6
Source DB: PubMed Journal: Biochim Biophys Acta ISSN: 0006-3002