Advances for treating in-hospital cardiac arrest: safety and effectiveness of a new automatic external cardioverter-defibrillator.

Antoni Martínez-Rubio1, Nabil Kanaan, Martin Borggrefe, Michael Block, Markku Mäkijärvi, Francesco Fedele, Carlo Pappone, Wilhelm Haverkamp, José Luis Merino, Gonzalo Barón Esquivias, Juan Cinca.   

Abstract

OBJECTIVES: The purpose of this study was to prospectively analyze the performance and safety of a new programmable, fully automatic external cardioverter-defibrillator (AECD) in a European multicenter trial. BACKGROUND Although, the response time to cardiac arrest (CA) is a major determinant of mortality and morbidity, in-hospital strategies have not significantly changed during the last 30 years.
METHODS: Patients (n = 117) at risk of CA in monitored wards (n = 51) and patients undergoing electrophysiologic testing or implantable cardioverter-defibrillator (ICD) implantation (n = 66) were enrolled. The accuracy of the automatic response of the device to any change of rhythm (lasting >1 s and >4 beats) was confirmed by reviewing the simultaneously recorded Holter data and the programmed parameters.
RESULTS: During 1,240 h, 1,988 episodes of rhythm changes were documented. A total of 115 episodes lasted > or =10 s or needed treatment (pacing, n = 32; ICD, n = 51; AECD, n = 35) for termination. The device detected ventricular tachyarrhythmias with a sensitivity of 100% and specificity of 97.6% (true negatives, n = 1,454; true positives, n = 499; false positives, n = 35; false negatives, n = 0). The false positives were all caused by T-wave oversensing during ventricular pacing. There were no complications or adverse events. The mean response time was 14.4 s for those episodes needing a full charge of the capacitor.
CONCLUSIONS: This new AECD is safe and effective in detecting, monitoring, and treating spontaneous arrhythmias. This fully automatic device shortens the response time to treatment, and it is likely that it will significantly improve the outcome of patients with in-hospital CA.

Entities:  

Mesh:

Year:  2003        PMID: 12598075     DOI: 10.1016/s0735-1097(02)02865-6

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  3 in total

Review 1.  Recent advances and controversies in adult cardiopulmonary resuscitation.

Authors:  Wanis H Ibrahim
Journal:  Postgrad Med J       Date:  2007-10       Impact factor: 2.401

2.  The automatic external cardioverter-defibrillator.

Authors:  Antoni Martínez-Rubio; Gonzalo Barón-Esquivias
Journal:  Indian Pacing Electrophysiol J       Date:  2004-07-01

3.  Automated external cardioversion defibrillation monitoring in cardiac arrest: a randomized trial.

Authors:  Bakhtiar Ali; Heather Bloom; Emir Veledar; Dorothy House; Robert Norvel; Samuel C Dudley; A Maziar Zafari
Journal:  Trials       Date:  2008-06-11       Impact factor: 2.279

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.