| Literature DB >> 12596988 |
K Akahane1, M Hashimoto, M Hada, Y Takano, K Ono, M Kai, T Kusama.
Abstract
As the survival rate of newborns has increased, the number of X ray computed tomography (CT) examinations performed on neonates has been increasing. The exposure doses from CT examinations are known to be higher than those from conventional radiography. Although radiation sensitivity of neonates is higher than that of adults, there are few reports on dose estimates of neonates in CT examinations. Four cylindrical phantoms and one neonatal phantom have been developed to estimate doses to neonates during CT examinations. Using these phantoms and glass dosemeters, absorbed doses were measured. Estimated exposure doses to neonates were higher than those to adults, and our results suggest a need to optimise carefully CT examinations in newborns.Mesh:
Year: 2003 PMID: 12596988 DOI: 10.1093/oxfordjournals.rpd.a006113
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972