Literature DB >> 12596633

Low background, pulsatile, in vitro flow circuit for modeling coronary implant thrombosis.

Kumaran Kolandaivelu1, Elazer R Edelman.   

Abstract

We have developed an in vitro method for creating pulsatile flows to mimic coronary type flow patterns on a beat-to-beat basis. The flow is created by accelerating fluid loops about an axis, inducing relative wall motion. Using this technique, a variety of oscillating flow patterns can be generated and modulated. Such flow generation offers the potential to monitor sensitive, flow-dependent, biological parameters like thrombosis while minimizing background disturbances from pump action and circuit effects. We examined this potential by measuring the loop occlusion time for loops stented with stainless steel 7-9 NIR stents and stentless control loops.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12596633     DOI: 10.1115/1.1517062

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  5 in total

1.  Stent thrombogenicity early in high-risk interventional settings is driven by stent design and deployment and protected by polymer-drug coatings.

Authors:  Kumaran Kolandaivelu; Rajesh Swaminathan; William J Gibson; Vijaya B Kolachalama; Kim-Lien Nguyen-Ehrenreich; Virginia L Giddings; Leslie Coleman; Gee K Wong; Elazer R Edelman
Journal:  Circulation       Date:  2011-03-21       Impact factor: 29.690

2.  Heparanase regulates thrombosis in vascular injury and stent-induced flow disturbance.

Authors:  Aaron B Baker; William J Gibson; Vijaya B Kolachalama; Mordechai Golomb; Laura Indolfi; Christopher Spruell; Eyal Zcharia; Israel Vlodavsky; Elazer R Edelman
Journal:  J Am Coll Cardiol       Date:  2012-04-24       Impact factor: 24.094

3.  A multichannel dampened flow system for studies on shear stress-mediated mechanotransduction.

Authors:  Peter L Voyvodic; Daniel Min; Aaron B Baker
Journal:  Lab Chip       Date:  2012-07-27       Impact factor: 6.799

4.  Quantification of thrombus formation in malapposed coronary stents deployed in vitro through imaging analysis.

Authors:  Jonathan Brown; Caroline C O'Brien; Augusto C Lopes; Kumaran Kolandaivelu; Elazer R Edelman
Journal:  J Biomech       Date:  2018-02-09       Impact factor: 2.712

Review 5.  Engineered arterial models to correlate blood flow to tissue biological response.

Authors:  Jordi Martorell; Pablo Santomá; José J Molins; Andrés A García-Granada; José A Bea; Elazer R Edelman; Mercedes Balcells
Journal:  Ann N Y Acad Sci       Date:  2012-04       Impact factor: 6.499

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.